REF: Alcolizer HH3 Alcohol Tester User Manual V6
Page 18
© Alcolizer Technology
Uncontrolled when printed
6
SPECIFICATIONS
The technical specifications of the HH3 are detailed in Table 1.
Table 1 - Alcolizer Technology HH3 Specifications
Instrument Application
Industry
–
Mouthpiece and Passive
BrAC Range
0.000 - 0.462 g/210L
Accuracy
Better than 0.005 at 0.100 g/210L BrAC
Test Time
Within 3-5 seconds at 0.100% BrAC
Recovery Time
Instantaneous for 0.0000 BrAC readings (reading remains on
screen for 10 seconds)
Operating Temperature
-5°C to +55°C
Operating Humidity
30% to 93% RH
Storage Temperature
-10°C to +70°C
Minimum Air Flow Rate
10L/min
Mouthpiece Type
Standard spigot locator or non-return mouthpiece.
(Mouthpieces are not required for Passive testing)
Sensor Type
Electro-chemical fuel cell (premium grade platinum)
Power Supply
Rechargeable Li-ION battery
Weight
272 grams (basic)
Dimensions
202mm H x 63mm W x 39mm D
Calibration
Six monthly recalibrations at Alcolizer’s service facility
7
ALCOCONNECT
AlcoCONNECT assists in the analysis of data downloaded from the HH3 instrument.
The AlcoCONNECT software can be downloaded from Alcolizer website
https://www.alcolizer.com/software-updates
onto a compatible computer
running MS Windows® XP, Vista, Win 7 or Win 8.
Installing Alcoconnect Software
To install the AlcoCONNECT software, proceed as follows:
•
Select AlcoCONNECT Toolbox Software on
https://www.alcolizer.com/software- updates
(Refer Figure 21).