REF: Alcolizer HH3 Alcohol Tester User Manual V6
Page 14
© Alcolizer Technology
Uncontrolled when printed
3.6.5
Taking a Passive Breath Sample
Perform a Passive Test as follows:
•
On the Start Mouthpiece Test page, press the down arrow key to access the Passive
Test function and Start Passive Test is displayed (refer Figure 16).
Figure 16 - Start Passive Test message
•
Press the Test button.
•
Hold the device no further than 100mm in front of the test subject’s mouth; so that
the subject is breathing or talking directly into the device’s sample inlet port (refer
Figure 17 - Providing a Passive sample
NOTE
If any step in a breath test sequence is not completed or the device is left idle for a
prolonged period of time, then the device will terminate the breath test and display
Test
timed out
. The device will return to the
Start, passive Test
display.
The standard breath test time out period is 60 seconds. If a test time out does occur
during a breath test, a new breath test must be performed. Press the Test key to perform
a new passive breath test as per steps above