REF: Alcolizer HH3 Alcohol Tester User Manual V6
Page 10
© Alcolizer Technology
Uncontrolled when printed
Attach the mouthpiece to the device as follows:
•
Grasp the mouthpiece at the non-ribbed end and tear the plastic wrapping from the
ribbed end, downwards to expose the sample inlet hole.
•
Locate the mouthpiece sample inlet port over the mouthpiece locator in line with
the recesses that surround the spigot (refer Figure 7).
•
Press the mouthpiece sample inlet port onto the locator until it is attached securely
(refer Figure 8). The orientation of the mouthpiece will not affect the sampling.
•
Remove the remaining portion of the plastic wrap just prior to testing to ensure
hygienic use.
•
A new mouthpiece must be used for each mouthpiece test.
Figure 7 - Locating the Mouthpiece
Figure 8 - Mouthpiece correctly positioned
3.6.2
Taking a Mouthpiece Breath Sample
Perform a Mouthpiece Test as follows:
NOTE
Before taking a breath sample, ensure the subject has not consumed alcohol for at least
15 minutes.
If any step in a breath test sequence is not completed or the device is left idle for a
prolonged period of time, then the device will terminate the breath test and display
Test
timed out
. The device will return to the
Start Mouthpiece Test
display
The standard breath test time out period is 60 seconds. If a test time out does occur
during a breath test, a new breath test must be performed.