1/3
E 40
Alcatel Vacuum Technology France - ASM 180 TD/TD+ - ASM 181 TD+ User’s Manual
Replacement / Recalibration
of the detector internal calibrated leak
Purpose of the
calibrated internal leak
Frequency of
internal leak
"recalibration"
How to recalibrate the
internal leak
Removal of the
internal calibrated leak
It enables the auto-
calibration of the detector.
Autocalibration is triggered
at start-up or when the
button is pressed on
the control panel (see
C 30
).
AUTOCAL
The frequency of preventive maintenance is listed
in D 10.
In order to ensure the reliability of the helium test,
ALCATEL recommends to "recalibrating" the internal calibrated
leak at least every 2 years (from the calibration date marked
on the leak label and its calibration certificate).
Recalibration is generally performed using a comparative
method with a reference standard. This work can only be
performed in ALCATEL or other approved service centers.
For this, therefore it is necessary to remove the internal
calibrated leak from the detector.
- Switch off the detector and disconnect it from the main
power.
- Open the front cover of the detector (attached with 4
screws).
- Disconnect the heat sensor connector (3-pin connector).
- Disconnect the DN 16 flange and remove the calibrated
leak.
Caution: Do not separate the heat sensor from the
calibrated leak.
Edition 03 - May 97
Calibrated leak
Components :
P/N
Internal calibrated leak . . . . . . . . . . . . . . 101302
Heat sensor
Label
DN 16