Alcatel Vacuum Technology France - ASM 180 TD/TD+ - ASM 181 TD+ User’s Manual
Cycle start faults
Edition 03 - May 97
S
YMPTOM
C
AUSE
R
EMEDY
Press the
filament key
,
the indicator light comes
on. Otherwise,
see D 60
.
NO
Release the
LDS key
to switch to vacuum test
mode
LDS mode selected
LDS key indicator
is ON
The cycle key is
disabled
(green
indicator light on)
The cycle key is
disabled
(indicator
lights off)
Switch off the detector
and
connect the I/O plug
connector. Start up again
(
see B 70
)
Note:
The cycle key green indicator light only indicates that
autocalibration is valided. The fact that the indicator light
is off does not inhibit the use of the cycle key.
Contact customer service
Filament off
filament key indicator
light is off
NO
Contact customer service
NO
I/O plug connector
This is not connected to the rear
of the detector