Configured Options
22 / 87
7001FAOPT73 ISS 13
General Options
1. Press the ON/OFF power button to the
OFF position.
2. Hold down the
START
button and press
the power button ON.
3. The main display will show “000”. Enter the
Access Code “
251
” using the rate selection
arrows.
4. Enter the first digit and press
NEXT
indicated on the screen.
5. When the complete code has been entered
press the
OK
softkey indicated on the
screen.
6. Select
GENERAL OPTIONS
using the
chevron keys and press the
OK
softkey
indicated on the screen.
7. Select the option you wish to
enable/disable or adjust and press the
MODIFY
softkey indicated on the screen.
8. When all the options that you require to
enable/disable or adjust have been
selected press the
OK
softkey indicated on
the screen.
9. When the set-up is complete, turn the
pump OFF and return the pump into
service, or select the next
configuration option.