Configured Options
23 / 87
7001FAOPT73 ISS 13
Fast Start and Backoff -
Enables drive slack fast start and occlusion auto bolus reduction.
VTBI -
Enables or disables the VTBI function.
Set VTBI over Time -
Sets a specific volume to be infused over a fixed period of time.
Induction -
Enables or disables the induction option.
Multidose -
Enables or disables the multidose option.
Rate Lock -
If enabled, the rate can be locked to prevent tampering of the set
infusion rate.
Quiet Mode -
If enabled, the low priority alarms and button pushes are muted.
Pressure Alarm -
Sets the pressure alarm value. Default is 300mmHg.
Max. Pressure Alarm -
Sets the maximum pressure limit. Default is 750mmHg.
Auto Pressure Alarm -
Enables or disables the automatic pressure alarm option.
AC Fail Alarm -
Enables or disables the AC Fail Alarm.
Rate Titration -
If enabled, the rate can be changed whilst the pump is infusing.
To titrate a rate while infusing, use the rate adjust keys to select
the new rate and press the START button to confirm the new rate.
NEOI Warning -
Sets the NEOI warning from 1 to 15mins.
EOI Point -
Sets the end of infusion volume as a percentage of the syringe
volume from 0.5% to 5.0%.
KVO at EOI -
If enabled the pump will switch to run at the KVO rate - 1.0ml/h
(or the current set rate if this is lower) at the EOI point. If disabled
the pump will stop at the EOI point.
Beam Alarm -
Sets up the default beam level between 0 and 7.
Weight -
Sets up the patient weight between 0.1kg and 150kg. Default is
70.0kg.
Purge Rate -
Sets the purge rate between 100 and 500ml/h. The upper limit
prevents excessive loss of fluid whilst purging.
Bolus -
Enables bolus delivery.
Default Bolus -
Sets the default bolus rate between 10 and1200ml/h.
Cap Bolus Rate -
Sets maximum permissible value for the bolus rate.
Cap Rate -
Sets maximum permissible value for the infusion rate.
Display Event Log -
If enabled the event log can be accessed via the options menu.
Log Last Patient Only -
If enabled the event log shows only the last patient (since new
patient selected). If disabled the complete log can be viewed.
Nurse Call -
Enables or disables the nurse call feature.
Nurse Call Inverted -
When enabled the nurse call feature is inverted.
Comms Pump Address -
Sets the Pump Communications Address.
Comms Monitor Only -
When disabled, allows remote control of the unit from the comms
link. Monitor is always possible.
Comms Odd Parity -
If enabled, communications parity is odd.
Comms ASCII -
If enabled, communications protocol is in ASCII.
General Options
The general options available are as follows: