Configured Options
24 / 87
7001FAOPT73 ISS 13
Clock Set
Clock Set is used to set the pumps internal clock.
1. Press the ON/OFF power button OFF.
2. Hold down the
START
button and press
the power button ON.
3. The main display will show “000”. Enter the
Access Code “
251
” using the rate selection
arrows.
4. Enter the first digit and press
NEXT
indicated on the screen.
5. When the complete code has been entered
press the
OK
softkey indicated on the
screen.
6. Select
CLOCK SET
using the chevrons
and press the
OK
softkey indicated on the
screen.
7. To adjust the value displayed use the
chevrons.
8. When the displayed value is correct press
the
NEXT
softkey to access the next field.
9. Repeat the process until the correct time
and date is displayed. Press the
OK
softkey to store the values.