10 / 87
Getting Started
7001FAOPT73 ISS 13
Review
REVIEW
1. To review the protocol data press the
REVIEW
softkey.
2. The protocol data will be displayed in the
main display.
3. Press the
REVIEW
softkey to toggle
between the pressure setting and the VTBI
status.
4. Repeat until the protocol data required is
displayed in the main display.
Pressure level with pressure set fitted (dedicated)
1. To check and adjust the pressure level
press the
PRESSURE
button. The display
will change and a bar graph will show the
pressure alarm level and current pressure
level.
2. To adjust the alarm level press the rate
keys to increase or decrease the level. The
new level will be indicated on the display.
3. When the pressure level has been selected
press
QUIT
to exit the screen.
4. If the Auto Pressure Option has been
enabled (see General Options) an
automatic pressure alarm level can be
calculated and set by pressing the
AUTO
softkey.
Pressure level without pressure set fitted (non-dedicated)
1. To check and adjust the pressure level
press the
PRESSURE
button. The display
will change and a bar graph will show the
pressure alarm level and current pressure
level.
2. To adjust the alarm level press the rate
keys to increase or decrease the level. The
new level will be indicated on the display.
3. When the pressure level has been selected
press
QUIT
to exit the screen.