Configured Options
21 / 87
7001FAOPT73 ISS 13
Drug Set-up
14. Select the Concentration Units using the
chevron keys, press the
OK
softkey to
select the concentration units.
15. Enter the Default Concentration of the drug
selected, using the chevron keys to
increase / decrease the default
concentration shown on the screen. When
the default concentration has been
selected press the
OK
softkey.
16. Select the Minimum Concentration of the
drug selected, using the chevron keys to
increase / decrease the minimum
concentration shown on the screen. When
the minimum concentration has been
selected press the
OK
softkey.
17. Select the Maximum Concentration of the
drug selected, using the chevron keys to
increase / decrease the maximum
concentration shown on the screen. When
the maximum concentration has been
selected press the
OK
softkey.
18. Select the Bolus Rate required, using the
chevron keys to increase / decrease the
bolus rate. When the rate has been
selected press the
OK
softkey.
19. Select the Maximum Bolus using the
chevron keys to increase / decrease the
volume. When the volume has been
selected press the
OK
softkey.
20. Select the Pressure Alarm level using the
chevron keys to increase / decrease the
level. When the level has been selected
press the
OK
softkey.
21. Review the drug setup data, press the
OK
softkey to confirm the data displayed. This
will return the pump to the drug setup
menu. Use the
BACK
softkey at any time
to return to the previous stage.