Getting Started
8 / 87
7001FAOPT73 ISS 13
Starting the Pump
1. Connect the pump to an AC power supply
using the AC power cable. Press the
ON/OFF button
2. NEW PATIENT? - Answering
NO
will retain
all previous patient history.
YES
will
automatically reset the patient history to
zero, press the
CONFIRM
softkey to
confirm selection. Check time and date is
correct.
3. CONFIRM SYRINGE - Check that the
syringe type and size being used matches
the display. If required, the type of syringe
can be changed by pressing the
TYPE
button.
4. Press
CONFIRM
when the correct type
and size are shown.
5. SET A RATE - Check the rate shown if old
patient data has been retained and change
the rate if necessary using the ADJUST
buttons.
6. PURGE (if required) - Press the
PURGE/BOLUS button then the two
PURGE softkeys together until fluid flows
and priming of the syringe extension line is
complete. The audible alarm will operate
during use of the
PURGE
softkeys and the
volume used during priming will be shown
on the display.
7. CONNECT
PATIENT
- Connect the
extension line to the patient access device,
and insert the pressure disc into the
pressure transducer.
8. START - Press START to commence
operation. INFUSING will be displayed.
The AMBER STOP light will be replaced by
the flashing GREEN START light to
indicate that the pump is operating.
Important:
The unit will automatically operate from its internal battery if the pump is
switched on without being connected to the AC power supply.
Each time the unit is switched ON, check that the alarm beeps twice and that all
the segments of the display, the green and amber lights are illuminated during
the self test routine.