RS232/Nursecall Feature
The RS232 / Nursecall feature fitted to this IVAC
®
“P” Series Syringe Pump allows the pump to be
monitored remotely and/or controlled via a suitable central monitoring or computer system.
When the pump is started by a command from the serial interface, communication must take
place over the serial interface, a communication must take place every 15 seconds or the pump
will alarm, display communications failure and stop infusing. This failure protects against failure
of the communications, including the removal of the RS232 cable.
Important:
The nursecall interface provides a remote backup to the internal audible alarm.
It should not be relied upon to replace monitoring of internal alarm.
Refer to the technical service manual for further information regarding the
RS232 interface. Since it is possible to control the syringe pump using the RS232
interface at some distance from the pump and hence remote from the patient,
responsibility for the control of the pump is vested in the software run on the
computer control system.
The assessment for the suitability of any software used in the clinical environment
to control or receive data from the pump lies with the user of the equipment. This
software should include detection of the disconnection or other failure of the
RS232 cable. The protocol is detailed in the technical service manual and is for
general information only. This relates to IVAC
®
“P” Series Syringe pumps fitted
with the RS232 communication interface.
Any connected analogue and digital components are required to meet EN60950
for data processing and EN60601 for medical devices. Anyone connecting
additional devices to the signal input or output is a system configurator and
responsible for meeting the requirements of the system standard EN60601-1-1.