Operating instructions

  

Digital ALMEMO

®

 D6 sensor

Atmospheric pressure sensor FDAD12SA

Temperature / humidity / atmospheric pressure sensor FHAD46x

Temperature / humidity / atmospheric pressure sensor FHAD36R

NTC psychrometer and atmospheric pressure sensor FNAD46

Infra-red sensor FIAD43

NTC  temperature sensor ZAD040FS / FS2

Hot-wire thermoanemometer and atmospheric pressure sensor FVAD35

Thermoanemometer FVAD05-TOKx

Rotating vanes FVAD15

Rotating vanes FVAD15H

Heat flow sensor FQAD00

CO

2

 and atmospheric pressure sensor FYAD00CO2

Precision pressure sensors FDAD33/35

Color temperature sensor FLAD23CCT

V-lambda-radiation sensor FLAD03VL1

V3.5

10.10.2018

www.ahlborn.com

Содержание ALMEMO D6

Страница 1: ...spheric pressure sensor FNAD46 Infra red sensor FIAD43 NTC temperature sensor ZAD040FS FS2 Hot wire thermoanemometer and atmospheric pressure sensor FVAD35 Thermoanemometer FVAD05 TOKx Rotating vanes...

Страница 2: ...ges 13 6 2 2 Technical data 13 7 D6 temperature humidity sensor FHAD46C 14 7 1 Measuring quantities and ranges factory default settings 14 7 2 Configuration on a PC via the sensor menu 15 7 2 1 Config...

Страница 3: ...our factory 30 14 2 Configuration on a PC via the sensor menu 30 14 2 1 Configurable measuring ranges 31 14 3 Technical data 31 15 D6 Thermo anemometer FVAD05 TOKx 32 15 1 Measuring ranges upon delive...

Страница 4: ...n on a PC via the sensor menu 42 20 2 1 Configurable measuring ranges 42 20 2 2 Measuring functions 43 20 3 Technical data 43 21 D6 color temperature sensor FLAD23CCT 44 21 1 Measuring ranges preset a...

Страница 5: ...O D6 sensor using measuring range DIGI supplies digital measured values from up to 4 measuring channels to the ALMEMO device where these are then processed as usual Any channel can be switched off dea...

Страница 6: ...f calibration the measuring accuracy will thus be deter mined no longer by the measuring instrument but exclusively by the sensor itself 2 3 Sensor menu Each D6 sensor has a stored individual sensor m...

Страница 7: ...e special D6 mea suring ranges for the D6 sensor and other settings At the interface the available mea suring ranges appear with new easy to understand abbreviations while on the measur ing instrument...

Страница 8: ...value Atmospheric pressure to compensate other sensors connected to the same ALMEMO device This programs abbreviation P in the designation of measuring channel D AP thus ensuring that this measured va...

Страница 9: ...ultisensormodul digital abgeglichen f r FHAD 46 C FH0D46C Multisensormodul digital abgeglichen f r FHAD 46 C2 FH0D46C2 ALMEMO D6 temperature humidity sensor FHAD 46 C pressure tight up to 16 bar FHAD4...

Страница 10: ...0 1 mb 5 2 Configuration on a PC via the sensor menu 5 2 1 Configurable measuring ranges Initially the ranges for the measuring channels can be configured from a list of two ranges factory default se...

Страница 11: ...C 5K Temperature 10 0 to 60 0 C Accuracy 2 K 0 to 60 C Refresh rate 1 second for all channels Connector colors 2 colors light gray and dark gray red lever Standard baud rate 115 2 kbaud freely selecta...

Страница 12: ...d in the ALMEMO measuring instrument as reference atmospheric pressure 6 1 Measuring ranges preset at our factory Description Range Exp Measuring range Units Resolution 1 Temperature T t B 01 DIGI 2 2...

Страница 13: ...on comprises the familiar abbreviations listed in tables issued by the Deutscher Wetterdienst German Meteorological Service and the newer symbols de fined in VDI VDE 3514 6 2 2 Technical data Operativ...

Страница 14: ...bles temperature relative humidity atmospheric pressure on the basis of formulae as per Dr Sonntag and the enhancement factor as per W B gel correction factor fw t p for real mixed gas systems The mea...

Страница 15: ...w 1 5 98 0 H 0 1 rH 3 Dewpoint DT td B 03 D td 1 C 0 1 K 4 Mixture MH r with PC B 04 D r 1 gk 0 1 g kg 5 Abs humidity AH dv B 05 D dv 1 gm 0 1 g m3 6 Vapor pressure VP e B 06 D e 1 mb 0 1 mb 7 Enthalp...

Страница 16: ...teresis typical 1 RH Atmospheric pressure 300 to 1100 mbar Accuracy 2 5 mbar 700 to 1100 mbar at 23 C 5 K Calculated quantities see 7 2 1 Atmospheric pressure compensation 0 to 6500 mbar programmable...

Страница 17: ...PC 8 1 Measuring ranges preset at our factory Description Range Exp Measuring range Units Resolution 1 Temperature T t B 01 DIGI 2 20 80 00 C 0 01 K 2 Rel Humidity RH Uw B 02 DIGI 1 5 98 0 H 0 1 rH 3...

Страница 18: ...ion comprises the familiar abbreviations listed in tables issued by the Deutscher Wetterdienst German Meteorological Service and the symbols more re cently defined in VDI VDE 3514 8 3 Technical data O...

Страница 19: ...anges factory default settings Designation Quantity Ex po nent Measuring range Units Resolution 1 Temperature T t B 01 DIGI 2 20 80 00 C 0 01 K 2 Rel humidity RH Uw B 02 DIGI 1 5 98 0 H 0 1 rH 3 Dewpo...

Страница 20: ...dienst German Meteorological Service and the symbols more recently defined in VDI VDE 3514 9 2 2 Technical data Operative range Temperature 20 to 80 Humidity 5 to 98 RH Measuring quantities and range...

Страница 21: ...re can also be used in the ALMEMO measuring instrument as reference atmospheric pressure see 3 3 1 10 1 Measuring ranges preset at our factory Description Range Exp Measuring range Units Resolution 1...

Страница 22: ...ignation comprises the familiar abbreviations listed in tables issued by the Deutscher Wetterdienst German Meteorological Service and the newer symbols de fined in VDI VDE 3514 10 3 Technical data Ope...

Страница 23: ...the basis of formulae as per Dr Sonntag and the enhancement factor as per W B gel correction factor fw t p for real mixed gas systems The measuring range and ac curacy of this system are thus much gre...

Страница 24: ...0 kJ 0 1 kJ kg The range the units 2 characters and a designation are programmed automatically this designation comprises the familiar abbreviations listed in tables issued by the Deutscher Wetterdien...

Страница 25: ...formulae with no additional error see 9 2 1 Atm pressure compensation 0 to 6500 mbar programmable Refresh rate 0 4 seconds for all four channels Connector colors 2 colors light gray and dark gray red...

Страница 26: ...ure to B 01 DIGI 1 40 0 600 0 C 0 1 C 12 2 Configuration on a PC via the sensor menu 12 2 1 Configurable measuring ranges Initially the ranges for the measuring channels can be configured from a list...

Страница 27: ...indow is being used these calculations may also have to take account of transmittance factory default 1 00 However this quantity can only be set in the sensor menu 12 3 Technical data Operative range...

Страница 28: ...0 C 0 01 K 2 Temperature T t Ntc2 Kl Ntc2 Gnd B 02 DNt2 2 50 125 00 C 0 01 K 13 2 Configuration on a PC via the sensor menu It is possible to activate not only the first but also as and when required...

Страница 29: ...te 0 3 seconds for 2 channels Connector colors 2 colors light gray and dark gray red lever Baud rate Standard 115 2 kbaud 1200 baud to 921 kbaud selectable Supply voltage 6 to 13 VDC Current consumpti...

Страница 30: ...Measuring ranges preset at our factory Description Range Exp Meas range Units Resolution 1 Temperature T t B 01 DIGI 1 20 70 00 C 0 1 K 2 Flow v with PC TH4 B 02 DIGI 3 0 08 2 000 m s 0 001 m s 2 Flow...

Страница 31: ...y 0 2 m s 2 of measured value 0 3 of measured value C 0 3 to 20 m s Response time 1 5 seconds Temperature compensation 0 to 50 C Temperature Measuring range 20 0 to 70 0 C Accuracy 0 7 C at 0 to 50 C...

Страница 32: ...ing range Dim Resolution 1 flow v 2 5 m s B 01 DIGI 3 0 050 to 2 500 m s 0 001 m s 2 atmospheric pressure AP p B 03 DIGI 1 300 to 1100 0 mbar 0 1 mbar The flow velocity of a thermo anemometer is inver...

Страница 33: ...k the enclosed sensor documentation 15 3 Technical data Flow Measuring range see under 15 2 1 Resolution 0 001 m s Accuracy 3 of measured value 1 of final value 2 digits Nominal temperature 23 C 2 K R...

Страница 34: ...serial form via RS485 driver To operate in sleep mode a 1 second wakeup delay is re quired 16 1 Measuring ranges preset at our factory Description Range Exp Meas range Units Resolution 1 example D6 S...

Страница 35: ...1 of final value 1 5 of measured value 0 4 20 00 m s 0 01 m s FVAD15 S140 1 of final value 1 5 of measured value 0 5 40 00 m s 0 01 m s FVAD15 S220 1 of final value 3 of measured value 0 6 20 00 m s...

Страница 36: ...e ALMEMO D6 plug When leaving our factory the ALME MO plug is preprogrammed with one measuring channel flow in m s In addition further measuring channels are available and can be selected via the sens...

Страница 37: ...0 00 to 135 00 m s 0 01 m s 9 D6 md20 v B 09 DIGI 2 0 00 to 22 50 m s 0 01 m s 10 D6 md40 v B 10 DIGI 2 0 00 to 45 00 m s 0 01 m s 11 D6 md80 v B 11 DIGI 2 0 00 to 90 00 m s 0 01 m s 12 D6 md120 v B 1...

Страница 38: ...should not be displayed it can be switched off deactivated and reactivated in the usual way via the ALMEMO device 18 2 Configuration on a PC via the sensor menu Initially the ranges for the four meas...

Страница 39: ...t incorporate its own temperature sensor the plate tem perature can also be entered manually in the sensor menu 18 3 Sensor connection The two sensors for heat flow mV and temperature NTC are clamped...

Страница 40: ...000 pp 1 ppm 2 Atm pressure B 02 DIGI 1 300 0 1100 0 mb 0 1 mb 19 2 Configuration on a PC via the sensor menu Initially the ranges for the measuring channels can be configured from a list factory defa...

Страница 41: ...ure compensation 0 to 6500 mbar programmable Current measuring time dCO2 15 seconds Total measuring time for averaging over 11 values DCO2 165 seconds Refresh rate 1 second for all channels Connector...

Страница 42: ...IGI 3 0 1 000 br 0 001 br The measuring range and resolution depend on the sensor type see data sheet 20 2 Configuration on a PC via the sensor menu Initially the ranges for the measuring channels can...

Страница 43: ...0 3 Technical data Measuring ranges Pressure depending on type see data sheet Resolution 0 002 full scale FS Accuracy 0 05 full scale FS 10 to 40 C 0 1 full scale FS 10 to 80 C Temperature 40 to 120 C...

Страница 44: ...in On a further sensor channel the illuminance can be obtained in lux lx or kilolux klx 21 1 Measuring ranges preset at our factory Description Range Exp Meas range Units Resolution 1 Color temperatur...

Страница 45: ...ange 1600 to 17000 K Coordinates resolution 0 005 Cosine correction 8 mm diffuser disc Cosine error 3 Measuring time 3 seconds Refresh rate 1 5 seconds for all channels Setting time 3 seconds Wakeup d...

Страница 46: ...range and three other channels with various resolutions for the lux range 22 1 Measuring ranges upon delivery Designation Range Exp Measuring range Dim Resolution 1 Ev kLux B 01 DIGI 2 0 to 200 00 kL...

Страница 47: ...ute Cos correction error f2 2 0 V lambda adaption 3 Linearity 1 Switch on time 1 s Switch off time 1 s Diffuser PTFE Weight approx 50g Measuring time 3 s Refresh rate 1 5 sec for all channels Setting...

Страница 48: ...3 83607 Holzkirchen Germany Tel 49 0 8024 3007 0 Fax 49 0 8024 300710 Internet http www ahlborn com email amr ahlborn com We take every possible care but the risk of inaccuracy can never be altogethe...