10. Measuring operations
10.1.2 Measuring ranges
With each channel switchover or sensor breakage the abbreviation for the
measuring range appears in the display. For identification purposes the follow-
ing table lists all possible measuring ranges supported by this device.
Transducer
Sensor / con-
nector
Measuring range Units Abbre-
viation
NiCr-Ni (K)
FT Axxx
-200.0...+1370.0
°C
NiCr
NiCroSil-NiSil (N)
ZA 9020-FSN
-200.0...+1300.0
°C
NiSi
Fe-CuNi (L)
ZA 9000-FSL
-200.0... +900.0
°C
FECO
Fe-CuNi (J)
ZA 9000-FSJ
-200.0... +950.0
°C
IrCo
Cu-CuNi (U)
ZA 9000-FSU
-200.0... +600.0
°C
CUCO
Cu-CuNi (T)
ZA 9000-FST
-200.0... +400.0
°C
CoCo
PtRh10-Pt (S)
FS Axxx
0.0...+1760.0
°C
Pt10
Ntc Typ N
FN Axxx
-20.00...+100.00
°C
Ntc
Millivolt
ZA 9000-FS0
-10.000...+55.000
mV
U 55
Millivolt 1
ZA 9000-FS1
-26.000...+26.000
mV
U 26
Millivolt 2
ZA 9000-FS2
-260.00...+260.00
mV
U260
Volt
ZA 9000-FS3
-0.2600...+2.6000
V
U2
.
60
Difference millivolt
ZA 9000-FS0D
-10.000...+55.000
mV
d 55
Difference millivolt 1
ZA 9000-FS1D
-26.000...+26.000
mV
d 26
Difference millivolt 2
ZA 9000-FS2D
-260.00...+260.00
mV
d260
Difference volt*
ZA 9000-FS3D
-0.2600...+2.6000
V
d2
.
60
Sensor voltage
any
0.00...20.00
V
UbAt
Milliampere
ZA 9601-FS1
0.000...+26.000
mA
I032
Percent (4 to 20mA)
ZA 9601-FS2
0.00... 100.00
%
P420
Frequency
ZA 9909-AK1
0... 32000
Hz
FrEq
Pulses
ZA 9909-AK2
0... 65000
PULS
Digital input
ZA 9000-EK2
0.0... 100.0
%
Inp
Digital inerface
ZA 9919-AKxx
-65000... +65000
diGi
Snap-on head normal 20
FV A915-S120
0.30... 20.00
m/s
S120
Snap-on head normal 40
FV A915-S140
0.40... 40.00
m/s
S140
Snap-on head micro 20
FV A915-S220
0.50... 20.00
m/s
S220
Snap-on head micro 40
FV A915-S240
0.60... 40.00
m/s
S240
Macro
FV A915-MA1
0.10... 20.00
m/s
L420
Water-Micro
FV A915-WM1
0.00... 5.00
m/s
L605
Relative air humidity, capacitive FH A646
0.0... 100.0
%H
°orH
Rel. air humidity, capacitive TC FH A646-C
0.0... 100.0
%H
HcrH
Rel. air humidity, capacitive TC FH A646-R
0.0... 100.0
%H
H rH
Mixture ratio, capacitive, with PC FH A646
0.0 ... 500.0
g/k
H AH
Dew-point temperature, cap.
FH A646
-25.0... 100.0
°C
H dt
Partial vapor pressure, cap.
FH A646
0.0 ...1050.0
mb
H UP
Enthalpy, capacitive
FH A646
0.0 ... 400.0
kJ
H En
* Double connectors with differential voltage / differential current (D-B) cannot be used
20 ALMEMO
®