Trouble-shooting  

necessary, insulate the sensor and eliminate interference by using
shielded or twisted wiring. 

Error: 

´

CALEr

´

is displayed when the device is switched on. 

Remedy: 

  The   calibration   of   a  measuring  range  may  have   become   misad-
justed. The device must be recalibrated at the factory. 

Error: 

Data transmission via the interface does not function. 

Remedy: 

Check interface module, connections, and settings. 
Are   both   devices   set   to   the   same   baud   rate   and   transmission
mode ? 
Is the correct COM interface on the computer being addressed ? 
To check the data flow and the handshake lines a small interface
tester with LEDs comes in very handy; (in ready-to-operate status
the data lines TXD, RXD carry negative potential of approx. -9V and
the LEDs light up green, whereas the handshake lines DSR, DTR,
RTS, CTS carry approx. +9V positive voltage and the LEDs light up
red.   For   the   duration   of   data  transmission   the   data  LEDs   should
flash red. 
Check   data   transmission   by   means   of   a   terminal   (AMR-Control,
WIN-Control, WINDOWS-Terminal). 
Address   the   device   using   its   assigned   device   number´

Gxy

´  (see

Manual 6.2.1). 
Enter <ctrl Q> for XON, if the device is in the XOFF status. 
Check the programming by means of ´

P15

´ (see Manual 6.2.3). 

Test   the  transmit   line   by  selecting  a   measuring  point  using  com-
mand ´

Mxx´

 and check in the display. 

Error: 

Data transmission in the network does not function.

Remedy: 

Check to ensure that all devices are set to different addresses. 
Address   all   devices   individually   via   the   terminal   using   command
´

Gxy

´. 

Addressed device is OK if at least ´

y CR

 

LF

´ is returned as echo. 

If transmission is still not possible, unplug the networked devices. 
Check   all  devices   individually  on  the  data  cable  to  the  computer;
(see above). 
Check the wiring for short-circuit or crossed wires. 
Are all network distributors supplied with power ? 
Network the devices again one at a time and check successively

If, after performing the above-listed checks and remedial steps, the device still
fails to behave as described in the operating instructions, it must be returned to
our factory in Holzkirchen, accompanied by an explanatory note, error descrip-
tion,  and if  available test  printouts.  With the AMR-Control software  you can
print out screen-shots with the relevant programming and save and / or print
out a comprehensive ´Function test´ in the device list or terminal operation. 

 ALMEMO

®

 2450-1

27  

Содержание ALMEMO 2450-1

Страница 1: ...____________________________ Operating instructions Universal measuring instrument ALMEMO 2450 1 V2 3 06 11 2013 www ahlborn com...

Страница 2: ...V not type L Mains adapter ZA1312 NA1 12V 0 2A Cable el isol ZA2690 UK 10 30V Interface RS 485 option 5 LCD Function field a Function b Measuring point 2nd meas value c Units for 2nd measured value Ma...

Страница 3: ...voltage monitoring 14 7 2 Mains operation 14 7 3 External DC voltage supply 14 7 4 Sensor supply 14 7 5 Switching ON OFF reinitialization 14 7 6 Data buffering 15 8 CONNECTING THE TRANSDUCERS 15 8 1...

Страница 4: ...ddress and networking 24 12 2 Analog output 25 12 3 Switching OFF automatically 26 12 4 Device locking 26 13 TROUBLE SHOOTING 26 14 DECLARATION OF CONFORMITY 28 15 APPENDIX 28 15 1 Technical data 28 1...

Страница 5: ...l ity tests We provide a guarantee lasting two years from delivery date that your device will function trouble free Before you send your device to us please observe the advisory notes in Chapter 13 Tr...

Страница 6: ...l of any such item as unsorted do mestic waste is strictly forbidden Please dispose of all packaging materials according to the applicable national waste management regulations Please dispose of cardb...

Страница 7: ...or conditions for which it is not suited Use of the device with an unsuitable power supply and or in conjunction with unsuitable peripheral equipment Use of the device for any purpose other than that...

Страница 8: ...uitable Be sure to observe the maximum load capacity of the sensor power supply Sensors with their own integrated power supply are not electrically iso lated from one another 4 2 Handling batteries re...

Страница 9: ...plete list of interface commands with all the printouts Manual Ch 7 The operating instructions you are now reading cover only those features and controls that are specific to this device Many sections...

Страница 10: ...t always indicate the correct units e g when a transmitter is connected Conversion between C Centigrade and F Fahrenheit is performed automatically Measured value designation Each sensor is identified...

Страница 11: ...most sensors sensor breakage is detected automatically except for con nectors with shunt dividers or additional electronics Analog output and scaling Each measuring point can be scaled by means of ana...

Страница 12: ...g over measuring point scans The measured values from measuring point scans can be averaged either over the whole measuring duration or over the specified cycle These average val ues can then be outpu...

Страница 13: ...ence value by pressing key CLR 6 Revert to normal measured value by pressing and holding down key CLR 7 Evaluating a measuring operation Call up maximum minimum values by pressing keys MAX MIN 6 To de...

Страница 14: ...d with power from an external source preferably using mains adapter ZA 1312 NA1 12 V 0 2 A via the DC socket 4 Please ensure that the mains voltage is correct 7 3 External DC voltage supply The DC soc...

Страница 15: ...isting sensors to ALMEMO instruments see Manual Ch 4 All suitable sensors with an ALMEMO connector see above have the measuring range and units al ready programmed and can thus be connected to the inp...

Страница 16: ...can flow between sensors power supply and peripherals All points must therefore lie at the same potential and or any un equal potentials must be electrically isolated Sensors combined within one conne...

Страница 17: ...digits b for depicting various measuring functions a there are also 2 arrows f for depicting the operating status Function field Main field Display of measuring functions in the function field Measur...

Страница 18: ...points Measuring range function not permitted Multiplexer M5 D B is not permitted Sensor breakage flashes Outside of measuring range undershoots cold junction compensation flashes or cold junction com...

Страница 19: ...Section 5 1 1 10 1 Measured value After switching ON first of all a segment test is performed then the battery voltage ap pears and if the batteries are almost empty 3 8 V the LOBAT arrow also appear...

Страница 20: ...olt 2 ZA 9000 FS2D 260 00 260 00 mV d260 Difference volt ZA 9000 FS3D 0 2600 2 6000 V d2 60 Sensor voltage any 0 00 20 00 V UbAt Milliampere ZA 9601 FS1 0 000 26 000 mA I032 Percent 4 to 20mA ZA 9601...

Страница 21: ...ature value can at the same time be dis played in the function field Select 2nd channel Activate temperature display Press and hold dow M Return to channel display Press and hold down M 10 2 Peak valu...

Страница 22: ...this channel The MAX MIN and MEM functions are thus also available for relative measurement 11 OUTPUTS The following interfaces corresponding accessories or options either are nec essary see 14 2 11 1...

Страница 23: ...e entered Or alternatively there are variants 2490 xR02 with 2 integrated electrically iso lated analog outputs see Section 15 2 these can be configured as 0 to 10 V 0 to 20 mA or 4 to 20 mA These app...

Страница 24: ...ling for Analog output P0 7 option see 12 2 Automatic switch OFF time in minutes see 12 3 To enter a value first press ON and the value starts flashing To modify the value press keys M or M To delete...

Страница 25: ...A for 30 0 to 120 0 C For the previously specified reference channel the values for analog output start and analog output end plus analog output type 0 to 20 mA or 4 to 20 mA can all be programmed Fir...

Страница 26: ...o grammed in many versatile ways It is suitable for connecting a wide variety of different sensors additional measuring instruments alarm signaling devices and peripheral equipment Given these numerou...

Страница 27: ...anual 6 2 1 Enter ctrl Q for XON if the device is in the XOFF status Check the programming by means of P15 see Manual 6 2 3 Test the transmit line by selecting a measuring point using com mand Mxx and...

Страница 28: ...owing advisory notes must be observed when operating the device Using the device in strong electromagnetic fields may aggravate measuring er rors 50 V at 3 V m and 1 5 meters thermocouple sensor After...

Страница 29: ...mm ABS acrylonitrile butadiene styrene weight approx 260 g Suitable conditions Operating temperature 10 to 50 C storage temperature 20 to 60 C Ambient relative humidity 10 to 90 rH non condensing 15...

Страница 30: ...conductor relays trigger input ZA 1006 EAK 15 3 Index Accessories 15 2 30 additional channels 8 2 15 AMR Control 5 1 3 13 analog output 11 2 23 Analog output 12 2 25 analog output end 12 2 25 analog o...

Страница 31: ...8 3 16 Power supply 15 1 14 29 Process control 5 1 3 12 Product overview 15 2 29 reference channel 12 2 25 reference value 10 4 22 reinitialization 7 5 14 Relative measuring 10 4 22 RS485 interface 11...

Страница 32: ...3 6 WIN Control 5 1 3 13 15 4 Your contact AHLBORN Mess und Regelungstechnik GmbH Eichenfeldstra e 1 83607 Holzkirchen Germany internet http www ahlborn com e mail amr ahlborn com We take every possib...

Отзывы: