Measured value
Transducer
Sensor / con-
nector
Measuring range Units Abbre-
viation
Function channels:
Maximum value of channel Mb1 any
Hi
Minimum value of channel Mb1
any
Lo
Aver.value M(t) over time of Mb1 any
A[t]
Aver. value M(n) of Mb2 to Mb1
any
A[n]
Sum S(n) of Mb2 to Mb1
any
S[n]
Total pulses S(t) of Mb1
ZA 9909-AK2
0... 65000
S[t]
Pulses / print cycle of Mb1
ZA 9909-AK2
0... 65000
S[P]
Alarm value of channel Mb1
any
Alrn
Measured value of Mb1
any
MEAS
Cold junction temperature
any
°C
CJ
Number of aver. values of Mb1
any
n(t)
Timer
any
s
tinE
TC=Temperature compensation PC=Pressure compensation
10.1.3 Double display
On all double sensors incorporating a tem-
perature sensor on the 1st channel the tem-
perature value can at the same time be dis-
played in the function field.
Select 2
nd
channel
Activate temperature display
Press and hold dow
M
▲
Return to channel display
Press and hold down
M
▲
10.2 Peak value memory
From the measured values acquired for each
measuring point the highest and the lowest
values are continuously recorded. To display
these high/low peak values first the desired
channel must be set (see Section 7.1) and
then the
MAX
or
MIN
key must be
pressed. As a check the display also includes
the associated symbol.
To display the maximum value press key :
MAX
.
To display the minimum value press key :
MIN
.
To delete the maximum value press and hold down key :
MAX
.
To delete the minimum value press and hold down key :
MIN
.
To return to the measuring point display press key :
M
▲
.
As soon as you clear the memory, the current measured value will appear (be -
cause measuring is continuous).
ALMEMO
®
2450-1
21