46
Agilent U1251A/U1252A User’s and Service Guide
2
Making
Measurements
In the range of 0–500
Ω
, the beeper will sound if the
resistance value falls below 10
Ω
. For other ranges, the
beeper will sound if the resistance falls below the typical
values indicated in the table below.
Conductance measurement eases measurement of very high
resistance of up to 100 G
Ω
. As high- resistance readings are
susceptible to noise, you can capture average readings via
the Dynamic Recording mode. See
Figure 17
on Page 59.
Measuring range
Beeper sounds when
500.00
Ω
< 10
Ω
5.0000 k
Ω
< 100
Ω
50.000 k
Ω
< 1 k
Ω
500.00 k
Ω
< 10 k
Ω
5.0000 M
Ω
< 100 k
Ω
50.000 M
Ω
< 1 M
Ω
500.00 M
Ω
< 10 M
Ω
Содержание U1251A
Страница 1: ...Agilent Technologies Agilent U1251A and U1252A Handheld Digital Multimeter User s and Service Guide...
Страница 34: ...34 Agilent U1251A U1252A User s and Service Guide 1 Getting Started Tutorial...
Страница 37: ...Making Measurements 2 Agilent U1251A U1252A User s and Service Guide 37 Figure 3 Measuring AC voltage MAX...
Страница 43: ...Making Measurements 2 Agilent U1251A U1252A User s and Service Guide 43 Figure 8 Measuring frequency Push MAX...
Страница 47: ...Making Measurements 2 Agilent U1251A U1252A User s and Service Guide 47 Figure 11 Conductance measurement MAX...
Страница 74: ...74 Agilent U1251A U1252A User s and Service Guide 3 Features and Functions Figure 26 Log review mode operation...
Страница 112: ...112 Agilent U1251A U1252A User s and Service Guide 5 Maintenance Figure 52 Battery charging procedures...
Страница 114: ...114 Agilent U1251A U1252A User s and Service Guide 5 Maintenance Figure 53 Fuse replacement Fuse 1 Fuse 2...
Страница 116: ...116 Agilent U1251A U1252A User s and Service Guide 5 Maintenance...
Страница 120: ...120 Agilent U1251A U1252A User s and Service Guide 6 Accessories...
Страница 156: ...156 Agilent U1251A U1252A User s and Service Guide 8 Specifications...