140
Agilent U1251A/U1252A User’s and Service Guide
7
Performance Tests and Calibration
To Change the Instrument Calibration Security Code from the
Front Panel
1
When the instrument in the unsecured mode, press
button for more than 1 second to enter Calibration
Security Code setting mode.
2
The factory default calibration security code 1234 will be
displayed on primary display.
3
Use the editing keys
and
to step each
character in the code.
4
Use the
and
to change each character in
the code.
5
Press
(Save) button, to store new calibration
security code.
6
If the new calibration security code has been successfully
store, the secondary display will show PASS.
Содержание U1251A
Страница 1: ...Agilent Technologies Agilent U1251A and U1252A Handheld Digital Multimeter User s and Service Guide...
Страница 34: ...34 Agilent U1251A U1252A User s and Service Guide 1 Getting Started Tutorial...
Страница 37: ...Making Measurements 2 Agilent U1251A U1252A User s and Service Guide 37 Figure 3 Measuring AC voltage MAX...
Страница 43: ...Making Measurements 2 Agilent U1251A U1252A User s and Service Guide 43 Figure 8 Measuring frequency Push MAX...
Страница 47: ...Making Measurements 2 Agilent U1251A U1252A User s and Service Guide 47 Figure 11 Conductance measurement MAX...
Страница 74: ...74 Agilent U1251A U1252A User s and Service Guide 3 Features and Functions Figure 26 Log review mode operation...
Страница 112: ...112 Agilent U1251A U1252A User s and Service Guide 5 Maintenance Figure 52 Battery charging procedures...
Страница 114: ...114 Agilent U1251A U1252A User s and Service Guide 5 Maintenance Figure 53 Fuse replacement Fuse 1 Fuse 2...
Страница 116: ...116 Agilent U1251A U1252A User s and Service Guide 5 Maintenance...
Страница 120: ...120 Agilent U1251A U1252A User s and Service Guide 6 Accessories...
Страница 156: ...156 Agilent U1251A U1252A User s and Service Guide 8 Specifications...