Specifications
8
Agilent U1251A/U1252A User’s and Service Guide
163
Temperature and Capacitance Specifications
[1] The accuracy is specified as following condition:
•
The accuracy does not include the tolerance of thermocouple probe. The
thermal sensor plugged into the meter should be placed in the operating
environment for at least an hour.
•
Use Null function to reduce the thermal effect. Before using NULL function,
set meter to no ambient compensation (
) mode and kept thermocouple
probe as close to the meter as possible, avoiding contact with any surface
that has a different temperature from ambient temperature.
•
When measuring temperature with respect to any temperature calibrator,
try to set both the calibrator and meter with external reference (without
internal ambient compensation). If both calibrator and meter are set with
internal reference (with internal ambient compensation), deviation may
show between the readings of the calibrator and meter, due to differences
in ambient compensation between calibrator and meter
.
[2] Only available in U1252A.
Function
Thermocouple
Type
Range Resolution
Accuracy
±
(% of r No. of
Least Significant Digit)
Temperature
(1)
K
-200 – 1372
°
C/
-328 – 2502
°
F
0.1
°
C/
0.1
°
F
0.3 % +3
°
C/
0.3 %+6
°
F
J
(2)
-210 – 1200
°
C/
-346 – 2192
°
F
0.1
°
C/
0.1
°
F
0.3 % +3
°
C/
0.3 %+6
°
F
Function
Range Resolution
Accuracy
±
(% of r Offset
Error)
Measuring Rate
At Full Scale
Max.
Display
Capacitance
10.000 nF
0.001 nF
1 %+8
4 times/sec.
11000
counts
100.00 nF
0.01 nF
1 %+5
1000.0 nF
0.1 nF
10.000
μ
F
0.001
μ
F
100.00
μ
F
0.01
μ
F
1000.0
μ
F
0.1
μ
F
1 time/sec.
10.000 mF
0.001 mF
0.1 times/sec.
100.00 mF
0.01 mF
3 %+10
0.01 times/sec.
Содержание U1251A
Страница 1: ...Agilent Technologies Agilent U1251A and U1252A Handheld Digital Multimeter User s and Service Guide...
Страница 34: ...34 Agilent U1251A U1252A User s and Service Guide 1 Getting Started Tutorial...
Страница 37: ...Making Measurements 2 Agilent U1251A U1252A User s and Service Guide 37 Figure 3 Measuring AC voltage MAX...
Страница 43: ...Making Measurements 2 Agilent U1251A U1252A User s and Service Guide 43 Figure 8 Measuring frequency Push MAX...
Страница 47: ...Making Measurements 2 Agilent U1251A U1252A User s and Service Guide 47 Figure 11 Conductance measurement MAX...
Страница 74: ...74 Agilent U1251A U1252A User s and Service Guide 3 Features and Functions Figure 26 Log review mode operation...
Страница 112: ...112 Agilent U1251A U1252A User s and Service Guide 5 Maintenance Figure 52 Battery charging procedures...
Страница 114: ...114 Agilent U1251A U1252A User s and Service Guide 5 Maintenance Figure 53 Fuse replacement Fuse 1 Fuse 2...
Страница 116: ...116 Agilent U1251A U1252A User s and Service Guide 5 Maintenance...
Страница 120: ...120 Agilent U1251A U1252A User s and Service Guide 6 Accessories...
Страница 156: ...156 Agilent U1251A U1252A User s and Service Guide 8 Specifications...