Features and Functions 3
Agilent U1251A/U1252A User’s and Service Guide
62
Figure 19 Refresh hold mode operation
N O T E
•
For voltage and current measurements, the holding value will not be
updated if the reading is below 500 counts.
•
For resistance and diode measurements, the holding value will not be
updated if the reading is in “
OL
” (open state).
•
The holding value may not be updated when the reading does not reach
the stable state for all measurements.
The
HOLD
sign flashes
until the measured
value is stable and
updated
Содержание U1251A
Страница 1: ...Agilent Technologies Agilent U1251A and U1252A Handheld Digital Multimeter User s and Service Guide...
Страница 34: ...34 Agilent U1251A U1252A User s and Service Guide 1 Getting Started Tutorial...
Страница 37: ...Making Measurements 2 Agilent U1251A U1252A User s and Service Guide 37 Figure 3 Measuring AC voltage MAX...
Страница 43: ...Making Measurements 2 Agilent U1251A U1252A User s and Service Guide 43 Figure 8 Measuring frequency Push MAX...
Страница 47: ...Making Measurements 2 Agilent U1251A U1252A User s and Service Guide 47 Figure 11 Conductance measurement MAX...
Страница 74: ...74 Agilent U1251A U1252A User s and Service Guide 3 Features and Functions Figure 26 Log review mode operation...
Страница 112: ...112 Agilent U1251A U1252A User s and Service Guide 5 Maintenance Figure 52 Battery charging procedures...
Страница 114: ...114 Agilent U1251A U1252A User s and Service Guide 5 Maintenance Figure 53 Fuse replacement Fuse 1 Fuse 2...
Страница 116: ...116 Agilent U1251A U1252A User s and Service Guide 5 Maintenance...
Страница 120: ...120 Agilent U1251A U1252A User s and Service Guide 6 Accessories...
Страница 156: ...156 Agilent U1251A U1252A User s and Service Guide 8 Specifications...