Programming High-Speed Test Extensions
Series N6700 User’s Guide
53
Front Panel:
SCPI Command:
Select
Transient\List\Pace
.
Select Dwell-paced. Press Select.
LIST:STEP AUTO, (@1)
In a trigger-paced list, the list advances one step for each trigger
received. To enable trigger-paced lists, select Trigger-paced on the
front panel menu. (Set the LIST:STEP command to ONCE.)
The dwell time associated with each step determines the minimum
time that the output remains at the step. If a trigger is received before
the dwell time completes, the trigger is ignored. To ensure that no
triggers are lost in a trigger-paced list, set the dwell time to zero.
Step 5.
Specify if you want the list to generate trigger signals that can be
used to trigger actions on other output channels or on external
equipment connected to the digital port.
Front Panel:
SCPI Command:
Select
Transient\List\Config
.
Select the List Step number.
To generate a trigger, enter a 1 in
the Tout Begin Step or Tout End
Step field. If a zero is entered, no
trigger is generated for the step.
Repeat this for each step. Use the
©
ª
keys to select the next step.
To program a trigger at the beginning
of step 4 for output 1, use
LIST:TOUT:BOST 0,0,0,0,1,0,
@(1)
To program a trigger at the end of
step 0, 2, and 4 for output 1, use
LIST:TOUT:EOST 1,0,1,0,1,0,
(@1)
Step 6.
Specify how you want the list to terminate.
Front Panel:
SCPI Command:
Select
Transient\List\Terminate
.
Select Stop Last Step. Press Select.
To program the output 1 list to remain
at the last list step when finished, use
LIST:TERM:LAST 1, (@1)
Step 7.
If applicable, specify how many times you want the list to repeat.
Sending the INFinity parameter in the SCPI command makes the list
repeat indefinitely. At reset, the list count is set to 1.
Front Panel:
SCPI Command:
Select the
Transient\List\Count
.
Enter a value in the Count field and
press Select.
To program the output 1 list to repeat
2 times, use
LIST:COUN 2, (@1)
Step 8.
Select a trigger source, initiate, and trigger the list. This is described
in detail under "Synchronizing Output Steps".
1
Trigger at BOST
0
2
3
4
5
1
Trigger at EOST
0
2
3
4
5