Connecting the Outputs
18
Series N6700 User’s Guide
Connecting the Outputs
WARNING
SHOCK HAZARD
Turn off AC power before making rear panel connections.
All wires and straps must be properly connected with the terminal block
screws securely tightened.
Disconnect the connector plug to make your wire connections. The
connector accepts wires sizes from AWG 12 to AWG 30. Note that
wire sizes smaller than AWG 20 are not recommended. Each
connector has four openings for attaching wires (see figure below).
Load connections are made at the + and - terminals. Sense
connections are made on the +s and -s terminals. Securely fasten the
wires by tightening the screw terminals.
After your wires are securely connected, insert the connector plug
into the back of the unit and secure it by tightening the locking
screws. A chassis ground binding post is available next to the AC
input connector for your convenience.
+S + -S
LOAD
TWIST LEADS
+
CONNECTOR
PLUG SHOWN
INSERT WIRES
TIGHTEN SCREWS
LOCKING SCREW
SENSE JUMPERS
INSTALLED FOR
LOCAL SENSING