Verification
128
Series N6700 User’s Guide
Test Record Form – Agilent N6731A
Agilent N6731A
Report No _______________
Date __________________
Description
Model
Minimum Specs.
Results
Maximum Specs.
Constant Voltage Tests
Voltage Programming & Readback
Minimum Voltage Vout
Front Panel Display Readback
High Voltage Vout
Front Panel Display Readback
+ 1 mV
Vout
−
20 mV
4.976 V
Vout
−
25 mV
__________
__________
__________
__________
+ 39 mV
Vout + 20 mV
5.024 V
Vout + 25 mV
CV Load Effect
−
2 mV
__________
+ 2 mV
CV Source Effect
−
1 mV
__________
+ 1 mV
CV Ripple and Noise
peak-to-peak
rms
N/A
N/A
__________
__________
+ 10 mV
+ 2 mV
Transient Response
Voltage at 200
µ
s
−
80 mV
__________
+ 80 mV
Constant Current Tests
Current Programming & Readback
Minimum Current Iout
Front Panel Display Readback
High Current Iout
Front Panel Display Readback
−
10 mA
Iout
−
20 mA
9.965 A
Iout
−
35 mA
__________
__________
__________
__________
+ 30 mA
Iout + 20 mA
10.035 A
Iout + 35 mA
CC Load Effect
−
2 mA
__________
+ 2 mA
CC Source Effect
−
1 mA
__________
+ 1 mA
Instrument
Settings
Test Description
N6731A
Voltage Programming & Readback, Zero Voltage
20 mV, 10 A
Voltage Programming & Readback, High Voltage
5 V, 10 A
CV Load Effect, Source Effect, Ripple and Noise
5 V, 10 A
Transient Response
5 V, from 5 A to 10 A
Current Programming & Readback, Zero Current
10 mA, 5 V
Current Programming & Readback, High Current
10 A, 5 V
CC Load Effect, Source Effect
10 A, 5 V