Calibration
Series N6700 User’s Guide
137
Front Panel:
SCPI Command:
Select
System\Cal\Sequences\CMRR
.
Then select Next.
CAL:VOLT:CMRR
(@1)
*OPC?
Current Programming and Measurement Calibration
Step 1.
Connect a precision shunt resistor to an output channel. Connect the
DVM to measure the voltage across the shunt. The shunt resistor
should be able to measure at least 120% of the channel’s rated full-
scale current.
Step 2.
Select the full-scale current programming range. The following
example selects the 10 amp full-scale current programming and
measurement range of channel 1. The value to program a range must
be the maximum current of the range.
Front Panel:
SCPI Command:
Select
System\Cal\Sequences\Iprog
Then select the High range.
CAL:CURR 10
, (@1)
Step 3.
Select the first current calibration point.
Front Panel:
SCPI Command:
Select Next. The information field
should indicate: Enter P1 data
CAL:LEV P1
*OPC?
Step 4.
Calculate the shunt current (I=V/R) and enter the data.
Front Panel:
SCPI Command:
Select the Measured Data field. Enter
the data from the external DVM.
Press Select when done.
CAL:DATA <data>
Step 5.
Select the second current calibration point.
Front Panel:
SCPI Command:
Select Next. The information field
should indicate: Enter P2 data
CAL:LEV P2
*OPC?
Step 6.
Calculate the shunt current (I=V/R) and enter the data.
Front Panel:
SCPI Command:
Select the Measured Data field. Enter
the data from the external DVM.
Press Select when done.
Press Next to finish calibration.
CAL:DATA <data>