G4208A User’s Guide
147
5
Maintenance and Repair
Firmware Updates
Upgrade from A.05.1x to B.xx
1
When the Agilent 1100/1200 series modules are on a firmware platform
A.05.09/1x, then update the modules to platform A.06.xx
2
Connect the Instant Pilot (with A.05.1x firmware) to one of the modules (if not
already done).
3
After boot, enter the
Maintenance
screen and press the
Controller
button.
Figure 101 Update Information for firmware A.05.11
4
Press the
Transfer
button. The Instant Pilot will switch into resident mode
(black screen) and the Update Tool will list the Instant Pilot.
5
Select the Instant Pilot for update to the target firmware revision
6
Start the update.
7
When finished, the Instant Pilot boots in main mode and shows all modules of
the system in the Welcome screen.
8
Close the connection and the LAN/RS- 232 Update Tool 2.10.