G4208A User’s Guide
133
4
Running Multiple-Vial Analyses
Synchronizing Analyses with External Devices
Wait for Multiple Start Request (Instant Pilot controls
injector)
After pressing the Start button, the Instant Pilot waits for an external start
requests before injection or sequence. The start requests have to be generated
by the external device. This mode is recommended when a device needs extra
time to get ready for the next analysis, and thus has to be in charge of the start
event. The receive of a Start Request leads first to a Prepeare (e.g. balance of a
detector) and afterwards to an injection by the sampler who then sends the Start
via APG remote.
Figure 90
Wait for repeated external start request