![Agilent Technologies InfinityLab LC Series Скачать руководство пользователя страница 109](http://html1.mh-extra.com/html/agilent-technologies/infinitylab-lc-series/infinitylab-lc-series_user-manual_2867765109.webp)
G4208A User’s Guide
109
3
Running an Isocratic Analysis
What You Will Need
What You Will Need
Instruments
Agilent Series isocratic, binary or quaternary pump, an autosampler and a
UV- detector.
Column
A 125 mm × 4.0 mm Hypersil ODS, 5
μ
m (Agilent Technologies part number
7982618- 564).
Solvents
For the isocratic pump, a solvent mixture of LC grade bidistilled water (35 %)
and acetonitrile (65 %).
Sample
The Agilent Technologies isocratic standard sample (Agilent Technologies
part number 01080- 68704). This contains 0.15 wt.% dimethylphthalate,
0.15 wt% diethylphthalate, 0.01 wt.% biphenyl and 0.03 wt.% o- terphenyl
dissolved in methanol.
NOTE
The descriptions are based on Agilent 1100/1200 systems. The Agilent 1290
Infinity System may have other requirements on the setup and material.