G4208A User’s Guide
128
4
Running Multiple-Vial Analyses
Re-calibrating With Multiple Groups of Standards
13
Change the settings according to
Figure 85
Sequence Calibration Wizard
14
You may press
Preview
to view the result.
15
Select
Done
to accept entries.
16
Press
Start
and confirm the start of the sequence.
The autosampler now analyzes:
•
three type A calibration standards in duplicate,
•
type A samples in vials 7 and 8,
•
three type A calibration standards in duplicate,
•
type A samples in vials 9 and 10,
•
three type A calibration standards in duplicate,
•
type A samples in vials 11 to 12,
•
three type A calibration standards in duplicate,
•
three type B calibration standards in duplicate,
•
type B samples in vials 13, 14 and 15,
•
three type B calibration standards in duplicate,
•
type B samples in vials 16,17 and 18,
•
three type B calibration standards in duplicate,
•
type B samples in vials 19, 20 and 21, and
•
three type B calibration standards in duplicate.