![Agilent Technologies Infinity Binary Pump VL 1290 Скачать руководство пользователя страница 85](http://html1.mh-extra.com/html/agilent-technologies/infinity-binary-pump-vl-1290/infinity-binary-pump-vl-1290_user-manual_2867763085.webp)
Agilent 1290 Infinity Binary Pump User Manual
85
Error Information
7
Pump Error Messages
Pump Error Messages
These errors are pump specific.
Pressure of binary pump above upper limit
Error ID: 22014
The pressure has exceeded the upper pressure limit.
•
Parameter: Measured pressure
Probable cause
Suggested actions
1
Blockage in flow path after the pressure
sensor.
•
Check for blockages in the LC system, e.g.
purge valve, Jet Weaver, degraded column,
column frits, needle, needle seat, capillaries
etc.
•
Check for particles in the solvent.
2
Inappropriate settings (pressure limit, flow
rate).
•
Decrease flow rate.
•
Increase pressure limit.
Содержание Infinity Binary Pump VL 1290
Страница 1: ...Agilent Technologies Agilent 1290 Infinity Binary Pump VL User Manual...
Страница 22: ...22 Agilent 1290 Infinity Binary Pump User Manual 2 Site Requirements and Specifications Specifications...
Страница 40: ...40 Agilent 1290 Infinity Binary Pump User Manual 3 Installing the Module Installation of seal wash option...
Страница 100: ...100 Agilent 1290 Infinity Binary Pump User Manual 7 Error Information Pump Error Messages...
Страница 108: ...108 Agilent 1290 Infinity Binary Pump User Manual 8 Test Functions and Calibrations Pump Head Leak Test...
Страница 188: ...188 Agilent 1290 Infinity Binary Pump User Manual 10 Parts and Materials for Maintenance Others...
Страница 202: ...202 Agilent 1290 Infinity Binary Pump User Manual 11 Identifying Cables Agilent 1200 Module to Printer...
Страница 222: ...222 Agilent 1290 Infinity Binary Pump User Manual 12 Hardware Information Early Maintenance Feedback...
Страница 267: ...Agilent 1290 Infinity Binary Pump User Manual 267 Index...