50
Agilent 1290 Infinity Binary Pump User Manual
4
Using the Pump
Setting up the Pump with the Instrument Control Interface
Method Parameter Settings
The Binary Pump method setup parameters are in eight sections:
• Flow
• Solvents A and B
• Stoptime
• Posttime
• Pressure Limits
• Timetable
• Advanced
• External Contacts
Table 5
Method parameters
Parameter
Limits
Description
Flow
0.00 – 5.00 mL/min
in steps of 0.001 .
Recommended flow
range: 0.05 –
5.00 mL/min .
The flow is the rate of movement of eluent along the column. It is
important that the flow rate is kept constant to ensure precise retention
time, and peak measurements. Variations in flow rate can occur as a result
of the partial failure of the pumping system, air in the pumping system, a
change in the mobile phase viscosity or a temperature change.
Solvents A and B
For each channel, you can select which of the two solvents to deliver. You
can set the percentage of solvent B to any value from 0 through 100 % .
Solvent A always delivers the remaining volume: 100 - %B. The solvent B
check boxes allow you to turn the solvent B channels on (checked) or off
(cleared).
When the
Use solvent types
check box in the
Compressibility
section is
on page 52), you click the down arrow
and select either a
Generic
solvent or a calibrated
Solvent
.
Stoptime
0.01 – 99999 min or
As Injector/No Limit
(an infinite run time).
The stoptime sets a time limit for your analysis. After the stoptime, all
gradients are stopped and the pump parameters return to their initial
values. The pump can be used as a stoptime master for the complete
analytical system. The pump also stops the detectors if they have a
No
Limit
stoptime setting. If no limit is given, a method will have to be stopped
manually.
Posttime
0.01 – 99999 min or
Off
(0.0 min ).
Your instrument remains in a not ready state during the posttime to delay
the start of the next analysis. You can use the
Posttime
to allow your
column to equilibrate after changes in solvent composition (for example
after gradient elution).
Содержание Infinity Binary Pump VL 1290
Страница 1: ...Agilent Technologies Agilent 1290 Infinity Binary Pump VL User Manual...
Страница 22: ...22 Agilent 1290 Infinity Binary Pump User Manual 2 Site Requirements and Specifications Specifications...
Страница 40: ...40 Agilent 1290 Infinity Binary Pump User Manual 3 Installing the Module Installation of seal wash option...
Страница 100: ...100 Agilent 1290 Infinity Binary Pump User Manual 7 Error Information Pump Error Messages...
Страница 108: ...108 Agilent 1290 Infinity Binary Pump User Manual 8 Test Functions and Calibrations Pump Head Leak Test...
Страница 188: ...188 Agilent 1290 Infinity Binary Pump User Manual 10 Parts and Materials for Maintenance Others...
Страница 202: ...202 Agilent 1290 Infinity Binary Pump User Manual 11 Identifying Cables Agilent 1200 Module to Printer...
Страница 222: ...222 Agilent 1290 Infinity Binary Pump User Manual 12 Hardware Information Early Maintenance Feedback...
Страница 267: ...Agilent 1290 Infinity Binary Pump User Manual 267 Index...