146
Agilent 1290 Infinity Binary Pump User Manual
9
Maintenance
Assembling the Pump Head
Assembling the Pump Head
See chapter
"Parts"
for details.
When
Before installing the pump head.
Tools required
p/n
Description
Pump head alignment tool
G4220-20012
Torque wrench 2 – 25 Nm
G4220-20013
4 mm hex bit
G4220-20014
2.5 mm Hex Bit
G4220-20015
Adapter ¼ in square to hex
01018-23702 Insert
tool
Parts required
p/n
Description
0905-1420
PE seals (pack of 2)
0905-1718
Wash Seal PE
5062-2484
Gasket, seal wash (pack of 6)
C A U T I O N
Limited life time of the pump head
Inserting the backup seal wrongly may limit the life time of the pump head.
➔
Please note the correct orientation of the backup seal.
C A U T I O N
Damage of the pump piston
The pump piston is very sensitive to shearing forces from the side.
➔
Use the alignment piston of the pump head alignment tool for the alignment
procedure described below.
Содержание Infinity Binary Pump VL 1290
Страница 1: ...Agilent Technologies Agilent 1290 Infinity Binary Pump VL User Manual...
Страница 22: ...22 Agilent 1290 Infinity Binary Pump User Manual 2 Site Requirements and Specifications Specifications...
Страница 40: ...40 Agilent 1290 Infinity Binary Pump User Manual 3 Installing the Module Installation of seal wash option...
Страница 100: ...100 Agilent 1290 Infinity Binary Pump User Manual 7 Error Information Pump Error Messages...
Страница 108: ...108 Agilent 1290 Infinity Binary Pump User Manual 8 Test Functions and Calibrations Pump Head Leak Test...
Страница 188: ...188 Agilent 1290 Infinity Binary Pump User Manual 10 Parts and Materials for Maintenance Others...
Страница 202: ...202 Agilent 1290 Infinity Binary Pump User Manual 11 Identifying Cables Agilent 1200 Module to Printer...
Страница 222: ...222 Agilent 1290 Infinity Binary Pump User Manual 12 Hardware Information Early Maintenance Feedback...
Страница 267: ...Agilent 1290 Infinity Binary Pump User Manual 267 Index...