![Agilent Technologies Infinity Binary Pump VL 1290 Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/agilent-technologies/infinity-binary-pump-vl-1290/infinity-binary-pump-vl-1290_user-manual_2867763049.webp)
Agilent 1290 Infinity Binary Pump User Manual
49
Using the Pump
4
Setting up the Pump with the Instrument Control Interface
Purge
Time
: 0 – 100.00 min
in steps of 0.01 .
Flow
: 0.000 –
5.000 mL/min for
each channel, in
steps of 0.001
(10.000 mL/min
maximum).
Setup and activation of
Purge
parameters. The automatic purge valve can
be used for purging the system. The process has been automated for ease
of use.
•
Off
: Turns off the purge.
•
On
: The device is purged.
•
Purge Flow
,
Time
and
Composition
during purge have to be defined.
As soon as the duration time of the purge ends, the module automatically
switches to analytical conditions again.
Prime
Select
On
to start priming,
Off
to turn priming off.
The
Prime
function is helpful for filling empty solvent lines or if air has
entered the pump heads. The module draws solvent, at high speed with all
four pump drives simultaneously, and dispenses it against the waste
position of the automatic purge valve. This is done 20 times, before the
process comes to an end.
Conditioning
at least 200 bar
(> 500 bar is better).
Use this function if you see excessive pressure or composition ripple, and
you are sure that the solvent type (aqueous/organic or specific
solvent/solvent mix) is correctly set, and there is no evidence of leakage in
the pump.
Conditioning may be necessary if the pump may contain air, for example
after running out of solvent, after a long period of standby or after service
or repair.
Table 4
Pump control parameters
Parameter
Limits
Description
Содержание Infinity Binary Pump VL 1290
Страница 1: ...Agilent Technologies Agilent 1290 Infinity Binary Pump VL User Manual...
Страница 22: ...22 Agilent 1290 Infinity Binary Pump User Manual 2 Site Requirements and Specifications Specifications...
Страница 40: ...40 Agilent 1290 Infinity Binary Pump User Manual 3 Installing the Module Installation of seal wash option...
Страница 100: ...100 Agilent 1290 Infinity Binary Pump User Manual 7 Error Information Pump Error Messages...
Страница 108: ...108 Agilent 1290 Infinity Binary Pump User Manual 8 Test Functions and Calibrations Pump Head Leak Test...
Страница 188: ...188 Agilent 1290 Infinity Binary Pump User Manual 10 Parts and Materials for Maintenance Others...
Страница 202: ...202 Agilent 1290 Infinity Binary Pump User Manual 11 Identifying Cables Agilent 1200 Module to Printer...
Страница 222: ...222 Agilent 1290 Infinity Binary Pump User Manual 12 Hardware Information Early Maintenance Feedback...
Страница 267: ...Agilent 1290 Infinity Binary Pump User Manual 267 Index...