![Agilent Technologies Infinity Binary Pump VL 1290 Скачать руководство пользователя страница 140](http://html1.mh-extra.com/html/agilent-technologies/infinity-binary-pump-vl-1290/infinity-binary-pump-vl-1290_user-manual_2867763140.webp)
140
Agilent 1290 Infinity Binary Pump User Manual
9
Maintenance
Disassembling the Primary Pump Head
Disassembling the Primary Pump Head
C A U T I O N
Damage of pump piston
The pump piston is made of ZrO
2
-based ceramic, which is a very hard and resistant
material, but it is sensitive to shearing forces from the side.
➔
Do not try to remove the pump piston from the rear.
➔
Do not use the piston for removing pump seals.
1
Remove the outlet valve at the top of the pump head and
the inlet valve at the bottom of the pump head.
2
For disassembling the pump head, remove the 4 hex
screws at the rear of the pump head.
Outlet valve
Inlet valve
Содержание Infinity Binary Pump VL 1290
Страница 1: ...Agilent Technologies Agilent 1290 Infinity Binary Pump VL User Manual...
Страница 22: ...22 Agilent 1290 Infinity Binary Pump User Manual 2 Site Requirements and Specifications Specifications...
Страница 40: ...40 Agilent 1290 Infinity Binary Pump User Manual 3 Installing the Module Installation of seal wash option...
Страница 100: ...100 Agilent 1290 Infinity Binary Pump User Manual 7 Error Information Pump Error Messages...
Страница 108: ...108 Agilent 1290 Infinity Binary Pump User Manual 8 Test Functions and Calibrations Pump Head Leak Test...
Страница 188: ...188 Agilent 1290 Infinity Binary Pump User Manual 10 Parts and Materials for Maintenance Others...
Страница 202: ...202 Agilent 1290 Infinity Binary Pump User Manual 11 Identifying Cables Agilent 1200 Module to Printer...
Страница 222: ...222 Agilent 1290 Infinity Binary Pump User Manual 12 Hardware Information Early Maintenance Feedback...
Страница 267: ...Agilent 1290 Infinity Binary Pump User Manual 267 Index...