Agilent 1290 Infinity Binary Pump User Manual
123
Maintenance
9
Replacing the Outlet valve
3
A gold seal between outlet valve and heat exchanger
capillary is used for a tight connection. The seal can be
replaced separately as needed.
4
Unscrew the outlet valve with a 14 mm wrench.
5
Insert the outlet valve into the pump head. Using a torque
wrench, set 10 Nm and close the outlet valve.
6
Insert the heat exchanger capillary into the outlet of the
outlet valve. Using a torque wrench with a 2.5 mm hex bit,
set 3 Nm and close the hex screw at the top of the outlet.
Hex screw
Gold seal
Heat exchanger
10 Nm
Содержание Infinity Binary Pump VL 1290
Страница 1: ...Agilent Technologies Agilent 1290 Infinity Binary Pump VL User Manual...
Страница 22: ...22 Agilent 1290 Infinity Binary Pump User Manual 2 Site Requirements and Specifications Specifications...
Страница 40: ...40 Agilent 1290 Infinity Binary Pump User Manual 3 Installing the Module Installation of seal wash option...
Страница 100: ...100 Agilent 1290 Infinity Binary Pump User Manual 7 Error Information Pump Error Messages...
Страница 108: ...108 Agilent 1290 Infinity Binary Pump User Manual 8 Test Functions and Calibrations Pump Head Leak Test...
Страница 188: ...188 Agilent 1290 Infinity Binary Pump User Manual 10 Parts and Materials for Maintenance Others...
Страница 202: ...202 Agilent 1290 Infinity Binary Pump User Manual 11 Identifying Cables Agilent 1200 Module to Printer...
Страница 222: ...222 Agilent 1290 Infinity Binary Pump User Manual 12 Hardware Information Early Maintenance Feedback...
Страница 267: ...Agilent 1290 Infinity Binary Pump User Manual 267 Index...