LAN Configuration
13
Initialization mode selection
Using Default
When
Using Default
is selected, the factory default parameters are taken
instead. These parameters enable a TCP/IP connection to the LAN interface
without further configuration, see
Figure 40
Using Default (Principle)
Since the default IP address is a so-called local address, it will not be routed by
any network device. Thus, the PC and the module must reside in the same
subnet.
The user may open a Telnet session using the default IP address and change
the parameters stored in the non-volatile memory of the module. He may then
close the session, select the initialization mode Using Stored, power-on again
and establish the TCP/IP connection using the new parameters.
When the module is wired to the PC directly (e.g. using a cross-over cable or a
local hub), separated from the local area network, the user may simply keep
the default parameters to establish the TCP/IP connection.
6Xi^kZ
EVgVbZiZg
9Z[Vjai
EVgVbZiZg
N O T E
Using the default address in your local area network may result in network problems. Take
care and change it to a valid address immediately.
Table 17
Using Default Parameters
IP address:
192.168.254.11
Subnet Mask:
255.255.255.0
Default Gateway
not specified
N O T E
In the
Using Default
mode, the parameters stored in the memory of the module are not
cleared automatically. If not changed by the user, they are still available, when switching
back to the mode Using Stored.
Содержание Infinity Binary Pump VL 1290
Страница 1: ...Agilent Technologies Agilent 1290 Infinity Binary Pump VL User Manual...
Страница 22: ...22 Agilent 1290 Infinity Binary Pump User Manual 2 Site Requirements and Specifications Specifications...
Страница 40: ...40 Agilent 1290 Infinity Binary Pump User Manual 3 Installing the Module Installation of seal wash option...
Страница 100: ...100 Agilent 1290 Infinity Binary Pump User Manual 7 Error Information Pump Error Messages...
Страница 108: ...108 Agilent 1290 Infinity Binary Pump User Manual 8 Test Functions and Calibrations Pump Head Leak Test...
Страница 188: ...188 Agilent 1290 Infinity Binary Pump User Manual 10 Parts and Materials for Maintenance Others...
Страница 202: ...202 Agilent 1290 Infinity Binary Pump User Manual 11 Identifying Cables Agilent 1200 Module to Printer...
Страница 222: ...222 Agilent 1290 Infinity Binary Pump User Manual 12 Hardware Information Early Maintenance Feedback...
Страница 267: ...Agilent 1290 Infinity Binary Pump User Manual 267 Index...