IDP-7 Dry Scroll Vacuum Pump 

DRAF

T 1/19/17

Installation

The IDP-7 is 

NOT

 intrinsically safe. Do not install or use

the pump in an environment exposed to atmospheric
agents (rain, snow, ice), dust, aggressive gases, or in
explosive environments or those with a high fire risk. 

During operation, the following environmental
conditions must be respected:

Connect the pump to the power supply using an IEC-320
style power cord of at least 10 A capacity (see Table 2 on
page 13).

Use

In order to reach maximum vacuum, the pump must be
left running for at least 24 hours with the inlet sealed. 

There are no special instructions for starting the pump; it
need only be switched on using the On/Off switch.

The optional inlet valve will open 30 seconds after the
pump is started.

There are no special instructions for stopping the pump; it
need only be disconnected from the electric power
source by the On/Off switch.

The optional inlet valve will close immediately after the
pump is stopped.

Maintenance

Personnel responsible for pump operation and mainte-
nance must be well-trained and aware of the accident
prevention rules.

If a pump is to be discarded, it must be disposed of in
accordance with specific national standards.

Disposal

Meaning of the 

WEEE

 logo found in labels.

The following symbol is applied in accordance with the
EC WEEE (Waste Electrical and Electronic Equipment)
Directive. This symbol (valid only in countries of the
European Community) indicates that the product it
applies to must NOT be disposed of together with ordi-
nary domestic or industrial waste but must be sent to a
differentiated waste collection system.

The end user is therefore invited to contact the supplier of
the device, whether the Parent Company or a retailer, to
initiate the collection and disposal process after checking
the contractual terms and conditions of sale.

For more information refer to: http://www.agilent.com/
environment/ product/ index.shtml.

Temperature:

+5 

°

C to +40 

°

C (41 

°

F to 104 

°

F)

Relative humidity:

0 to 95% (non-condensing)

CAUTION

When installing the vacuum pump it is 
recommended that the pump be posi-
tioned in a way that the power switch 
and power cord connection remain eas-
ily accessible to the user in order to be 
available as an option to disconnect 
power to the pump quickly if needed.

WARNING

The pump is designed for operation 
with neutral or noncorrosive fluids. It 
is absolutely forbidden to use it with 
potentially explosive or inflammable 
substances.

WARNING

Death may result from contact with 

high voltages. Always take extreme care 
and observe the accident prevention 
regulations in force.

When machine is powered up, 

be careful of moving parts and high 
voltages. 

If you have to perform maintenance 

on the pump after a considerable time 
in operation, allow it to cool as the tem-
perature of the outer surface may be in 
excess of 60 

°

C.

Always disconnect your power 

supply to the pump before beginning 
maintenance work.

NOTE

Before returning the pump to the factory 
for repair, the “Request for Return” 
sheet attached to this instruction man-
ual must be completed and sent to the 
local sales office. A copy of the sheet 
must be inserted in the pump package 
before shipping.

Содержание IDP-7

Страница 1: ...D R A F T 1 1 9 1 7 Vacuum Products Division INSTRUCTION MANUAL IDP 7 Dry Scroll Vacuum Pump Manual No X3807 90009 Revision A January 2017...

Страница 2: ...D R A F T 1 1 9 1 7 Copyright 2017 Agilent Inc IDP 7 Dry Scroll Vacuum Pump Slo Blo and Littelfuse are registered trademarks of Littelfuse Inc...

Страница 3: ...luding representatives of Seller which are incon sistent or in conflict with the terms of this warranty shall not be binding upon Seller unless reduced to writing and approved by an officer of Seller...

Страница 4: ...nd Inspection 11 Installation 12 Safety 12 Startup 12 Electrical Connections 12 Start Current 14 Mechanical Connections 15 Operation 16 Cleaning the Pump 16 Startup Procedure 16 Shutdown Procedure 16...

Страница 5: ...and storing the pump the following environmental requirements should not be exceeded Preparation for Installation The pump is supplied in a special protective packing If this shows signs of damage whi...

Страница 6: ...irable operation NOTE The equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC rules These limits are designed to provide reasonabl...

Страница 7: ...must NOT be disposed of together with ordi nary domestic or industrial waste but must be sent to a differentiated waste collection system The end user is therefore invited to contact the supplier of...

Страница 8: ...erference that may cause undesired oper ation If this equipment does cause harmful interference to radio or television reception which can be determined by turn ing the equipment off and on the user i...

Страница 9: ...ences environnementales suivantes ne doivent pas tre d pass es Pr paration pour l installation La pompe est livr e dans un emballage protecteur sp cial Si elle manifeste des signes de dommages qui peu...

Страница 10: ...quer une op ration ind sirable REMARQUE L quipement a t test et d clar con forme aux limites d un dispositif num rique de Classe A selon la Partie 15 des r gles de la FCC Ces limites sont destin es f...

Страница 11: ...oit seulement tre allum e en utilisant le commutateur On Off La vanne d admission en option s ouvrira 30 secondes apr s le d marrage de la pompe Il n existe aucune instruction sp ciale pour l arr t de...

Страница 12: ...nisseur du dispositif maison m re ou revendeur pour mettre en uvre le processus de collecte et mise au rebut Pour en savoir plus consulter http www agilent com environment product index shtml AVERTIS...

Страница 13: ...AC below 230 VAC Ultimate pressure 20 mT at 60 Hz 30 mT at 50 Hz Inlet connection NW25 Exhaust connection NW16 Gas ballast Female 1 8 National Pipe Thread 20 micron sintered plug provided SAE 2 5 16 2...

Страница 14: ...IDP 7 Dry Scroll Vacuum Pump 10 D R A F T 1 1 9 1 7 Figure 1 Outline Drawing with Dimensions Figure 2 Pumping Speed Curves...

Страница 15: ...Cowling 9 Mounting Holes 4 through holes for 1 4 or M6 Hardware 10 Rear Cowling 11 KA40 Phillips PT hardware 6 12 On Off Switch 13 Fuse Holder and Voltage Changeover Tab 14 Power Connection IEC 320 15...

Страница 16: ...ire Install in an area that is not exposed to rain steam or excessive humidity They can cause electric shock short circuits and severe bodily injury Before reconfiguring the pump voltage or inspecting...

Страница 17: ...rent your product will draw An under sized cord will cause a drop in line voltage resulting in loss of power and overheating Table 2 IDP 7 Power Cord Selection Country Power Cord Specification Order E...

Страница 18: ...7 Start Current Table 3 lists the typical IDP 7 maximum starting current and duration Table 3 Typical IDP 7 Maximum Starting Current A Frequency High Voltage Range 100 V 120 V 127 V 200 V 230 V 240 V...

Страница 19: ...d to normal ambient humidity Gas ballast port 2 is used for high water vapor loads such as repetitively pumping out a volume exposed to excessive ambient humidity or pro cess gas loads with a high wat...

Страница 20: ...configured for the mains voltage to which the pump is connected See Electri cal Connections on page 12 Figure 6 ON OFF Power Switch 2 Switch the pump ON 3 If the pump is equipped with an optional inl...

Страница 21: ...llast plugged Replace breather vent Contact Agilent Solvent in pump Flush pump with air or dry nitrogen Install trap or filter Seals worn out Rebuild pump Poor conductance to pump Replumb with shorter...

Страница 22: ...mp periodically at atmosphere for a minute or two to flush it out For more information see Cleaning the Pump on page 16 Tip Seal Replacement The parts and tools required to replace tip seals include T...

Страница 23: ...Tip Seal onto the Orbiting Scroll item 9 3 Sequentially insert the seal from center to the outer edge of the scroll wall 4 Cut the Tip Seal about 1 8 3 mm from the groove end 5 Use the remaining Tip S...

Страница 24: ...e 2 Disassemble and repair as necessary The pump is now ready to return to service 1 Front Cowling Bolts M8 3 2 Exhaust Fitting NW16 3 Front Cowling 4 Frame Bolts M5 4 5 Scroll Housing 6 Main O ring P...

Страница 25: ...he core product e g inlet screens vent valves o Prior to shipment and if applicable for your product drain any oils or other liquids purge or flush all gasses and wipe off any excess residue o If orde...

Страница 26: ...15 business days Failure to do so or returning a non rebuildable part crashed will result in an invoice for the non returned non rebuildable part o Returns for credit toward the purchase of new or ref...

Страница 27: ...e Chemical name Chemical Symbol CAS Number Toxic YES NO Harmful YES NO Corrosive YES NO Reactive YES NO Flammable YES NO Explosive YES NO Radioactive YES NO Biological YES NO Oxidizing YES NO Sensitiz...

Отзывы: