![Agilent Technologies IDP-7 Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/agilent-technologies/idp-7/idp-7_instruction-manual_2868768017.webp)
13
IDP-7 Dry Scroll Vacuum Pump
DRAF
T 1/19/17
Power Cord
Several power cord options are available from your
Agilent dealer. Descriptions of the available power cords
and their ordering numbers are given in Table 2.
Grounding Instructions
This product must be grounded. In the event of an electri-
cal short circuit, grounding reduces the risk of electric
shock by providing an escape wire for the electric cur-
rent. This pump is equipped with a power cord that has a
grounding wire with an appropriate grounding plug. The
plug must be inserted into an outlet that is properly
installed and grounded in accordance with all local codes
and ordinances.
Extension Cords
If you must use an extension cord with this product:
❑
For this product, Agilent recommends using only
extension cords with a minimum of 16-gage wire and
a maximum length of 25 feet (7.6 m).
❑
Use only a 3-wire extension cord that will accept the
plug.
❑
Make sure your extension cord is in good condition.
❑
Be sure the extension cord is rated high enough to
carry the current your product will draw. An under-
sized cord will cause a drop in line voltage resulting in
loss of power and overheating.
Table 2 IDP-7 Power Cord Selection
Country
Power Cord Specification
Order
Europe
10 A / 220-230 VAC, 2.5 m 656494220
Denmark
10 A / 220-230 VAC, 2.5 m 656494225
Switzerland
10 A / 230 VAC, 2.5 m
656494235
UK/Ireland
13 A / 230 VAC, 2.5 m
656494250
India
10 A / 220-250 VAC, 2.5 m 656494245
Israel
10 A / 230 VAC, 2.5 m
656494230
Japan
12 A / 100 VAC, 2.3 m
656494240
North
America
15 A / 125 VAC, 2.0 m
656458203
10 A / 230 VAC, 2.5 m
656494255
DANGER
Improper installation of the grounding
plug can result in a risk of electrical
shock.
WARNING
Check with a qualified electrician or
serviceman if the grounding instructions
are not completely understood, or if
you are in doubt as to whether the
product is properly grounded.
Do not modify the plug provided; if it
does not fit the outlet, have the proper
outlet installed by a qualified electrician.
Connect the product only to an outlet
that has the same configuration as the
plug.
Do not use an adapter with this
product.
WARNING
If the product must be reconnected
for use on a different type of electric cir-
cuit, the connector should be replaced
by qualified service personnel.