Agilent Technologies IDP-7 Скачать руководство пользователя страница 5

1

IDP-7 Dry Scroll Vacuum Pump 

DRAF

T 1/19/17

Instructions for Use

General Information

This equipment is designed for use by trained personnel.
The user should read this instruction manual and any
other additional information supplied by Agilent before
operating the equipment. Agilent will not be held respon-
sible for any events that occur due to non-compliance
with these instructions, improper use by untrained per-
sons, non-authorized interference with the equipment, or
any action contrary to that provided for by specific
national standards.

The IDP-7 is a hermetic, dry scroll vacuum pump. This
pump is suitable for pumping air or inert gases. The pump
is not intended to pump corrosive, explosive, or particu-
late-forming gases.

The following paragraphs contain all the information nec-
essary to guarantee the safety of the operator when using
the equipment. Detailed information is supplied in
“Technical Information” on page  9.

This manual uses the following standard safety protocol:

Storage

When transporting and storing the pump, the following
environmental requirements should not be exceeded:

 

Preparation for Installation

The pump is supplied in a special protective packing. If
this shows signs of damage, which may have occurred
during transport, contact your local sales office.

Total weight of the packing, IDP-7 pump included, is
approximately 28.1 kg (62 lbs). 

WARNING

The warning messages are for attracting 
the attention of the operator to a partic-
ular procedure or practice which, if not 
followed correctly, could lead to serious 
injury.

CAUTION

The caution messages are displayed 
before procedures, which if not fol-
lowed, could cause damage to the 
equipment.

NOTE

The notes contain important informa-
tion taken from the text.

Temperature:

20 

°

C to +60 

°

C (

°

F to 140 

°

F)

Relative humidity:

0 to 95% (non-condensing)

WARNING

 The IDP-7 pump itself weighs 24.5 kg 
(54 lbs). To avoid injury, use lifting aids 
and proper lifting techniques when 
moving the pump.

WARNING

When unpacking the pump, be sure not 
to drop it and avoid any kind of sudden 
impact or shock vibration to it.

NOTE

Normal exposure to the environment 
cannot damage the pump. Neverthe-
less, it is advisable to keep the pump 
inlet closed until the pump is installed 
in the system.

Содержание IDP-7

Страница 1: ...D R A F T 1 1 9 1 7 Vacuum Products Division INSTRUCTION MANUAL IDP 7 Dry Scroll Vacuum Pump Manual No X3807 90009 Revision A January 2017...

Страница 2: ...D R A F T 1 1 9 1 7 Copyright 2017 Agilent Inc IDP 7 Dry Scroll Vacuum Pump Slo Blo and Littelfuse are registered trademarks of Littelfuse Inc...

Страница 3: ...luding representatives of Seller which are incon sistent or in conflict with the terms of this warranty shall not be binding upon Seller unless reduced to writing and approved by an officer of Seller...

Страница 4: ...nd Inspection 11 Installation 12 Safety 12 Startup 12 Electrical Connections 12 Start Current 14 Mechanical Connections 15 Operation 16 Cleaning the Pump 16 Startup Procedure 16 Shutdown Procedure 16...

Страница 5: ...and storing the pump the following environmental requirements should not be exceeded Preparation for Installation The pump is supplied in a special protective packing If this shows signs of damage whi...

Страница 6: ...irable operation NOTE The equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC rules These limits are designed to provide reasonabl...

Страница 7: ...must NOT be disposed of together with ordi nary domestic or industrial waste but must be sent to a differentiated waste collection system The end user is therefore invited to contact the supplier of...

Страница 8: ...erference that may cause undesired oper ation If this equipment does cause harmful interference to radio or television reception which can be determined by turn ing the equipment off and on the user i...

Страница 9: ...ences environnementales suivantes ne doivent pas tre d pass es Pr paration pour l installation La pompe est livr e dans un emballage protecteur sp cial Si elle manifeste des signes de dommages qui peu...

Страница 10: ...quer une op ration ind sirable REMARQUE L quipement a t test et d clar con forme aux limites d un dispositif num rique de Classe A selon la Partie 15 des r gles de la FCC Ces limites sont destin es f...

Страница 11: ...oit seulement tre allum e en utilisant le commutateur On Off La vanne d admission en option s ouvrira 30 secondes apr s le d marrage de la pompe Il n existe aucune instruction sp ciale pour l arr t de...

Страница 12: ...nisseur du dispositif maison m re ou revendeur pour mettre en uvre le processus de collecte et mise au rebut Pour en savoir plus consulter http www agilent com environment product index shtml AVERTIS...

Страница 13: ...AC below 230 VAC Ultimate pressure 20 mT at 60 Hz 30 mT at 50 Hz Inlet connection NW25 Exhaust connection NW16 Gas ballast Female 1 8 National Pipe Thread 20 micron sintered plug provided SAE 2 5 16 2...

Страница 14: ...IDP 7 Dry Scroll Vacuum Pump 10 D R A F T 1 1 9 1 7 Figure 1 Outline Drawing with Dimensions Figure 2 Pumping Speed Curves...

Страница 15: ...Cowling 9 Mounting Holes 4 through holes for 1 4 or M6 Hardware 10 Rear Cowling 11 KA40 Phillips PT hardware 6 12 On Off Switch 13 Fuse Holder and Voltage Changeover Tab 14 Power Connection IEC 320 15...

Страница 16: ...ire Install in an area that is not exposed to rain steam or excessive humidity They can cause electric shock short circuits and severe bodily injury Before reconfiguring the pump voltage or inspecting...

Страница 17: ...rent your product will draw An under sized cord will cause a drop in line voltage resulting in loss of power and overheating Table 2 IDP 7 Power Cord Selection Country Power Cord Specification Order E...

Страница 18: ...7 Start Current Table 3 lists the typical IDP 7 maximum starting current and duration Table 3 Typical IDP 7 Maximum Starting Current A Frequency High Voltage Range 100 V 120 V 127 V 200 V 230 V 240 V...

Страница 19: ...d to normal ambient humidity Gas ballast port 2 is used for high water vapor loads such as repetitively pumping out a volume exposed to excessive ambient humidity or pro cess gas loads with a high wat...

Страница 20: ...configured for the mains voltage to which the pump is connected See Electri cal Connections on page 12 Figure 6 ON OFF Power Switch 2 Switch the pump ON 3 If the pump is equipped with an optional inl...

Страница 21: ...llast plugged Replace breather vent Contact Agilent Solvent in pump Flush pump with air or dry nitrogen Install trap or filter Seals worn out Rebuild pump Poor conductance to pump Replumb with shorter...

Страница 22: ...mp periodically at atmosphere for a minute or two to flush it out For more information see Cleaning the Pump on page 16 Tip Seal Replacement The parts and tools required to replace tip seals include T...

Страница 23: ...Tip Seal onto the Orbiting Scroll item 9 3 Sequentially insert the seal from center to the outer edge of the scroll wall 4 Cut the Tip Seal about 1 8 3 mm from the groove end 5 Use the remaining Tip S...

Страница 24: ...e 2 Disassemble and repair as necessary The pump is now ready to return to service 1 Front Cowling Bolts M8 3 2 Exhaust Fitting NW16 3 Front Cowling 4 Frame Bolts M5 4 5 Scroll Housing 6 Main O ring P...

Страница 25: ...he core product e g inlet screens vent valves o Prior to shipment and if applicable for your product drain any oils or other liquids purge or flush all gasses and wipe off any excess residue o If orde...

Страница 26: ...15 business days Failure to do so or returning a non rebuildable part crashed will result in an invoice for the non returned non rebuildable part o Returns for credit toward the purchase of new or ref...

Страница 27: ...e Chemical name Chemical Symbol CAS Number Toxic YES NO Harmful YES NO Corrosive YES NO Reactive YES NO Flammable YES NO Explosive YES NO Radioactive YES NO Biological YES NO Oxidizing YES NO Sensitiz...

Отзывы: