Agilent InfinityLab LC Series Multicolumn Thermostats User Manual
37
2
Site Requirements and Specifications
Performance Specifications
Column identifica
-
tion
Optionally, column identification kit to track history of up to four
columns.
Mounted on the left hand-side of module.
Instrument control
LC and CE Drivers A.02.14 or above.
Agilent Instant Pilot (G4208A) B.02.20 or above.
InfinityLab LC Companion (G7108A) 1.1 or above.
Lab Advisor B.02.08 or above.
For details about supported
software versions refer to the
compatibility matrix of your ver
-
sion of the LC & CE Drivers
Communication
Controller Area Network (CAN)
G7116A is a hosted module.
(The LC stack needs to contain
suitable host module or a LAN
card for communication and
control).
Maintenance and
safety-related fea
-
tures
Extensive diagnostics, error detection and display (through Instant
Pilot control module and Agilent Lab Advisor), leak detection, safe
leak handling, leak output signal for shutdown of pumping system.
Low voltages in major maintenance areas.
GLP features
Valve heads carrying tags with serial number, pressure rating,
number of switches and valve type.
Concept of column identification.
Housing
All materials recyclable.
1
All specifications are valid for distilled water at ambient temperature (25 °C), set point at 40 °C and a stable flow range from 0.2 - 5
mL/min. Equilibration time: 10 min.
Table 4
Performance Specifications G7116A
Type
Specification
Comments