Agilent InfinityLab LC Series Multicolumn Thermostats User Manual
115
7
Error Information
MCT Error Messages
WRONG VALVE
Error ID: 32130
Valve not supported in this module.
Probable cause
Suggested actions
1
A valve head has been installed which is not
supported by G7116A.
Only valve heads up to 800 bar or with up to four
positions can be used in G7116A. Exchange the
valve head with a suitable one. For changing the
valve head follow the instructions
Replace Valve
Heads
.
NOTE
For G7116B:
•
G4232D (5067-4240: 2-position/10-port Valve 1300 bar)
•
G4231C (5067-4241: 2-position/6-port Valve 1300 bar)
•
G4234A (5067-4284: 6-position/14-port Valve 800 bar)
•
G4234C (5067-4273: 6-position/14-port Valve 1300 bar)
•
G4239C (5067-4233: 8-position/18-port Valve 1300 bar)
For G7116B Bio:
•
G5641A (5067-6682: 2-position/10-port Valve 1300 bar, bio)
For G7116A:
•
G4232C (5067-4283: 2-position/10-port Valve 800 bar)
•
G4231A (5067-4282: 2-position/6-port Valve 800 bar)
•
G4237A (5067-4279: 4-position/10-port Valve 800 bar)
For G7116A Bio-inert:
•
G5632A (5067-4132: 2-position/10-port Valve 600 bar, bio-inert)
•
G5631A (5067-4148: 2-position/6-port Valve 600 bar, bio-inert)
•
G5639A (5067-4134: 4-position/10-port Valve 600 bar, bio-inert)