Agilent InfinityLab LC Series Multicolumn Thermostats User Manual
32
2
Site Requirements and Specifications
Physical Specifications
Physical Specifications
Physical Specifications G7116B
Table 1
Physical Specifications G7116B
Type
Specification
Comments
Weight
12.5 kg (27.6 lbs)
Dimensions
(height × width × depth)
160 x 435 x 436 mm
(6.3 x 17.1 x 17.2 inches)
Width with column identification kit:
472 mm
G7116B may have a
Column ID tag reader on
both sides.
Line voltage
100 – 240 V~, ± 10 %
Wide-ranging capability
Line frequency
50 or 60 Hz, ± 5 %
Power consumption
150 VA, 150 W
Ambient operating tempera
-
ture
4–55 °C (39–131 °F)
Ambient non-operating tem
-
perature
-40 – 70 °C (-40 – 158 °F)
Humidity
< 95 % r.h. at 40 °C (104 °F)
Non-condensing
Operating altitude
Up to 3000 m (9842 ft)
Safety standards:
IEC, EN, CSA, UL
Overvoltage category II, Pollution degree 2 For indoor use only.
ISM Classification
ISM Group 1 Class B
According to CISPR 11