76
S:\agilent\e8285\USRGUIDE\BOOK\CHAPTERS\opoverv.fb
Chapter 3, Operating Overview
Measurement Triggering Process
Default Trigger Modes
Default Trigger Mode for Remote Operation: CDMA and Analog Measurements
Upon powerup, or upon receiving a *RST command, or upon pressing the front
panel Preset key, the Test Set’s default trigger mode for remote operation of both
CDMA and analog measurements is Repetitive
retriggering
with Full settling.
If a local-to-remote transition occurs, the trigger mode will be set to the last
trigger mode command received while in remote mode. If no trigger mode
command has been received since powerup, the default state is set.
Default Trigger Mode for Local Operation: CDMA Measurements
Upon powerup, or upon receiving a *RST command, or upon pressing the front
panel Preset key, the Test Set’s default trigger mode for front panel operation of
the FER and Rho Suite of Measurements is Single retriggering.
If a remote-to-local transition occurs, the trigger mode will be determined by:
•
FER -
the state of the
Meas Cntl
field on the
CDMA CELLULAR MOBILE
RECEIVER TEST screen
•
Rho Suite of Measurements
-
the state of the
Meas Cntl
field on the
CDMA
CELLULAR MOBILE TRANSMITTER TEST screen
NOTE:
Setting the
Meas Cntl
field to
Cont
(Continuous) sets the trigger mode to repetitive
retriggering.
Upon powerup, or upon receiving a *RST command, or upon pressing the front
panel Preset key, or upon a remote-to-local transition the Test Set’s default trigger
mode for front panel operation of the Avg/Chan Power measurement is Repetitive
retriggering.
Default Trigger Mode for Local Operation: Analog Measurements
Upon powerup, or upon receiving a *RST command, or upon pressing the front
panel Preset key, the Test Set’s default trigger mode for front panel operation of
analog measurements is Repetitive
retriggering
with Full settling.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание E8285A
Страница 18: ...Contents 17 Index Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 100: ...99 4 Status Reporting Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 268: ...267 6 IBASIC Controller Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 346: ...345 A Error Messages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...