364
S:\agilent\e8285\USRGUIDE\BOOK\CHAPTERS\errorms.fb
Error Messages
Error
−
222
Data out of range
Indicates that a legal program data element was parsed but could not be executed because
the interpreted value was outside the legal range as defined by the device (see IEEE 488.2,
11.5.1.1.5).
Error
−
223
Too much data
Indicates that a legal program data element of block, expression, or string type was
received that contained more data than the device could handle due to memory or related
device- specific requirements.
Error
−
224
Illegal parameter value
Used where exact value, from a list of possibles, was expected.
Error
−
230
Data corrupt or stale
Possibly invalid data; new reading started but not completed since last access.
Error
−
231
Data questionable
Indicates that measurement accuracy is suspect.
Error
−
240
Hardware error
Indicates that a legal program command or query could not be executed because of a
hardware problem in the device.
Error
−
241
Hardware missing
Indicates that a legal program command or query could not be executed because of
missing device hardware. For example, an option was not installed.
Error
−
250
Mass storage error
Indicates that a mass storage error occurred.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание E8285A
Страница 18: ...Contents 17 Index Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 100: ...99 4 Status Reporting Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 268: ...267 6 IBASIC Controller Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 346: ...345 A Error Messages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...