101
Chapter 4, Status Reporting
Status Reporting
Status Reporting
This section describes the status reporting structure used in the Test Set. The
structure is based on the IEEE 488.1-1987 and 488.2-1987 Standards and the
Standard Commands for Programmable Instruments (SCPI) Version 1994.0.
For a full listing of command mnemonics, refer to the “Status”
syntax
,
in GPIB
Command Syntax chapter of the Agilent Technologies E8285A Condensed
Programming Reference Guide.
Status Register Structure Overview
The structure of the register groups used in the Test Set is based upon the status
data structures outlined in the IEEE 488 and SCPI 1994.0 Standards. There are
two types of status data structures used in the Test Set: status registers and status
queues. The general models, components, and operation of each type of status
data structure are explained in the following sections.
Status Register Model
This section explains how the status registers are structured in the Test Set. The
generalized status register group model shown on the previous page consists of a
Condition Register, Transition Filters, an Event Register, an Enable Register, and
a Summary Message Bit.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание E8285A
Страница 18: ...Contents 17 Index Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 100: ...99 4 Status Reporting Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 268: ...267 6 IBASIC Controller Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 346: ...345 A Error Messages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...