Chapter 2, Configuring Your Test Set
General Operating Information
To Set RF Voltage Interpretation (50 ohm/emf)
1. Access the CONFIGURE screen.
2. Position the cursor in front of the
RFGen Volts
field.
3. Press the CURSOR CONTROL knob or press the Enter key to select
50 ohm
or
emf
.
Voltage settings can control either:
•
the voltage across a 50-ohm load, or
•
the open circuit voltage (emf).
This setting affects the RF Generator and Tracking Generator amplitudes.