51
Chapter 3, Operating Overview
To Change the Measurement Display
To Set a Measurement Reference
The REF SET function establishes a measurement reference point. This allows
you to make a direct comparison between two measurement results, or between a
measurement standard and the actual measurement results.
Referenced measurements are displayed as either a ratio (dB) or difference
between the measured value and the reference.
Front-Panel Example
1. Position the cursor in front of the unit-of-measure for the measurement you want to set
the reference for.
2. Press and release the Shift key, then the INCR
÷
10 key to select the REF SET function.
3. Enter the reference value.
4.
Ref
appears below the measurement value to indicate that a reference has been set. The
measurement field may display a different unit-of-measure, and limit choices for units.
GPIB Example
The following GPIB command also sets a 10 dBm reference for
Avg Power
measurements.
“DISP CCNT;MEAS:CDM:AVGP:REF 10 DBM”