366
S:\agilent\e8285\USRGUIDE\BOOK\CHAPTERS\errorms.fb
Error Messages
Error
−
260
Expression error
Indicates that an expression program data element related error occurred.
Error
−
261
Math error in expression
Indicates that a syntactically legal expression program data element could not be executed
due to a math error. For example, a divide-by-zero was attempted.
Error
−
270
Macro error
Indicates that a macro-related execution error occurred.
Error
−
271
Macro syntax error
Indicates that a syntactically legal macro program data sequence, according to IEEE
488.2, 10.7.2, could not be executed due to a syntax error within the macro definition (see
IEEE 488.2, 10.7.6.3).
Error
−
272
Macro execution error
Indicates that a syntactically legal macro program data sequence could not be executed
due to some error in the macro definition (see IEEE 488.2, 10.7.6.3).
Error
−
273
Illegal macro label
Indicates that the macro label defined in the *DMC command was a legal string syntax,
but could not be accepted by the device (see IEEE 488.2, 10.7.3 and 10.7.6.2). For
example, the label was too long, the same as a common command header, or contained
invalid header syntax.
Error
−
274
Macro parameter error
Indicates that the macro definition improperly used a macro parameter placeholder (see
IEEE 488.2, 10.7.3).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание E8285A
Страница 18: ...Contents 17 Index Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 100: ...99 4 Status Reporting Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 268: ...267 6 IBASIC Controller Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 346: ...345 A Error Messages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...