278
S:\agilent\e8285\USRGUIDE\BOOK\CHAPTERS\ibasic.fb
Chapter 6, IBASIC Controller
Interfacing to the IBASIC Controller using Serial Ports
Receive and Transmit Pacing
When receiving characters into the
IBASIC Command Line
field, the Test Set’s
microprocessor responds to each entry and no buffering is required. Therefore,
when using your PC or terminal to send characters to the
IBASIC Command
Line
field, it is permissible to set
Flow Cntl
to
None
.
When sending data through the Test Set’s Serial Port to external devices like
printers which may have small input buffers, it is important to set
RFlow Cntl
to
Xon/Xoff
. This allows the printer to stop data transmission from the Test Set
when the printer’s buffer is full and then start it again when the printer is ready.
The Test Set has a Serial Port input buffer length of 2000 characters. Buffer size
becomes important when IBASIC programs expect to receive large amounts of
data through the Serial Port with a single ENTER statement.
Table 33
Test Set Serial Port 9 Configuration
Field
Available Settings
Recommended Setting
Serial In
Inst/IBASIC
Inst
IBASIC Echo
On/Off
On
Inst Echo
On/Off
On
Serial Baud Rate
150, 300, 600, 1200, 2400, 4800,
9600, 19200
9,600
Parity
None, Odd, Even, Always 1,
Always 0
None
Data Length
7 bits, 8 bits
8 bits
Stop Length
1 bit, 2 bits
1 bit
Flow Cntl
None, Xon/Xoff, Hardware
Xon/Xoff
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание E8285A
Страница 18: ...Contents 17 Index Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 100: ...99 4 Status Reporting Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 268: ...267 6 IBASIC Controller Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 346: ...345 A Error Messages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...