328
7
SCPI Command Reference
:SENSe:UDConverter:HARMonic
Equivalent key
TR Menu -> Time Offset -> Wide Span
TR Menu -> Span -> Wide Span
:SENSe:UDConverter:HARMonic
Syntax
:SENSe:UDConverter:HARMonic <numeric>
:SENSe:UDConverter:HARMonic?
Description
Sets/Reads the factor of the frequency offset
Parameter
Equivalent key
PN Menu -> System -> Instrument Setup -> Frequency Offset (User Downconv.) ->
Harmonic #
SP Menu -> System -> Instrument Setup -> Frequency Offset (User Downconv.) ->
Harmonic #
FP Menu -> System -> Instrument Setup -> Frequency Offset (User Downconv.) ->
Harmonic #
TR Menu -> System -> Instrument Setup -> Frequency Offset (User Downconv.) ->
Harmonic #
USER Menu -> System -> Instrument Setup -> Frequency Offset (User Downconv.) ->
Harmonic #
:SENSe:UDConverter:LO
Syntax
:SENSe:UDConverter:LO <numeric>
:SENSe:UDConverter:LO?
Description
Sets/Reads the LO frequency of the frequency offset
Maximum value which can be set is 100/Harmonic (GHz).
Parameter
When the conversion mode used in the frequency offset is RF=N*LO+IF (Allowable
Unit
s
Resolution
10n
<Numeric>
<Description>
Range
1 to 20
Preset value
1
Unit
-
Resolution
1
Содержание E5052A
Страница 4: ...4 ...
Страница 28: ...28 1 Making Effective Use of This Manual How to Use This Manual ...
Страница 51: ...3 51 Making a Measurement Trigger System 3 Making a Measurement Figure 3 1 Trigger system ...
Страница 58: ...58 3 Making a Measurement Waiting for End of measurement ...
Страница 76: ...76 4 Reading Writing Measurement Data Entering Data in a Trace ...
Страница 82: ...82 Chapter5 Saving and Recalling File Management Managing Files ...
Страница 476: ...476 AppendixB Status Reporting System Status Register Structure Figure B 3 Status register structure 1 of 6 ...
Страница 478: ...478 AppendixB Status Reporting System Status Register Structure Figure B 5 Status register structure 3 of 6 ...
Страница 480: ...480 AppendixB Status Reporting System Status Register Structure Figure B 7 Status register structure 5 of 6 ...