272
7
SCPI Command Reference
:FORMat:DATA
Parameter
Equivalent key
No equivalent key is available on the front panel.
:FORMat:DATA
Syntax
:FORMat:DATA {ASCii|REAL32|REAL64}
:FORMat:DATA?
Description
Sets/Reads data transfer mode
Parameter
Equivalent key
No equivalent key is available on the front panel.
:HCOPy:ABORt
Syntax
:HCOPy:ABORt
Description
Aborts printing (No Query)
Equivalent key
PN Menu -> System -> Abort Printing
SP Menu -> System -> Abort Printing
FP Menu -> System -> Abort Printing
TR Menu -> System -> Abort Printing
USER Menu -> System -> Abort Printing
:HCOPy:IMAGe
Syntax
:HCOPy:IMAGe {NORMal|INVert}
:HCOPy:IMAGe?
Description
NORMal(Preset
value)
Set byte order so that a byte containing MSB (Most Significant
Bit) is transferred first
SWAPped
Set byte order so that a byte containing LSB (Least Significant
Bit) is transferred first
Description
ASCii(Preset
value)
Set data transfer mode to 'ASCii'
REAL32
Set data transfer mode to 'IEEE 32 floating point binay'
REAL64
Set data transfer mode to 'IEEE 64 floating point binay'
Содержание E5052A
Страница 4: ...4 ...
Страница 28: ...28 1 Making Effective Use of This Manual How to Use This Manual ...
Страница 51: ...3 51 Making a Measurement Trigger System 3 Making a Measurement Figure 3 1 Trigger system ...
Страница 58: ...58 3 Making a Measurement Waiting for End of measurement ...
Страница 76: ...76 4 Reading Writing Measurement Data Entering Data in a Trace ...
Страница 82: ...82 Chapter5 Saving and Recalling File Management Managing Files ...
Страница 476: ...476 AppendixB Status Reporting System Status Register Structure Figure B 3 Status register structure 1 of 6 ...
Страница 478: ...478 AppendixB Status Reporting System Status Register Structure Figure B 5 Status register structure 3 of 6 ...
Страница 480: ...480 AppendixB Status Reporting System Status Register Structure Figure B 7 Status register structure 5 of 6 ...